ZEISS
2 Requirements | 2.3 Mains Connection
Service
Access Area
1000
500
2500
3600
800
1000
1300
1200
400
500
5
7
4
3
2
max. 1500
6
1
Fig. 2: Recommended arrangement for Axio Observer 7 (measurements in mm)
1
LSM scan module
2
Component rack with laser module and
power supply unit
3
System table, medium
4
Microscope stand Axio Observer 7
5
Power point for accessories
6
Gas supplies (if required)
7
Customer arranged computer
workspace
The laser module, the power supply and, if available, the Airyscan 2 for the scanning module are
mounted inside the component rack, which should be positioned next to the system table on the
floor.
2.3 Mains Connection
The Microscope System must be properly installed with the power cords and supplied socket with
protective ground contact securely connected.
The provided electrical connection must be in accordance with the applicable electrical codes for
the country of installation. In order to avoid disturbance from other installed machines, you must
use a separate power connection.
The Microscope System is connected to the AC mains via two mains cables, one running to a
power strip and one running to the Power Supply unit for further power distribution.
§
Make sure the mains connector is not more than 1.5 m away from the position of the laser
rack.
§
Do not extend or modify the supplied power cords.
§
Do not connect electrical systems that are not authorised by ZEISS on the supplied power sup-
ply cords.
Value
Nominal AC voltage
1/N/PE 120 or 230 VAC (+/-10%)
2 standard local power points are required.
Nominal frequency
50 - 60 Hz
Installation Requirements ZEISS LSM 900 | en-US | Rev. 2 | 349500-8183-000
9