LSM 880
Level-adjustable Sample Holder for LSM
ZEISS
10/2014 V_01
000000-2071-464
21
9
Level-adjustable Sample Holder for LSM
Caution! Adjustable stage insert screws may intersect with parts of the microscope, e.g. the
condenser unit. When moving a motorized stage, avoid any collisions. Do not install inserts
while the stage is moving. Be careful when switching the condenser front lens.
9.1
Stage Leveling Using LSM in Reflection Mode and Focus Drive
Fig. 9
Adjustable sample holder – schematic
view
Fig. 10
Focus control with focus drive. Course of
reflection on tilted (A) and even (B)
coverslip when moving focus upwards
(left to right)
0.
Adjust the edges of the clefts viewed from the right side (Fig. 9/
csv
) and back side (Fig. 9/
cbv
) to be
parallel by using screws
sy
and
sx
, respectively (Fig. 9). The cleft’s width should be approximately
2 mm as preset by the factory, but can be adjusted with screw
sz
if needed.
1.
First locate the area of interest of your sample. It is important to level the sample in the same region
as the actual measurements, as the coverslip may be bent or tilted at other locations.
2.
Prepare a LSM reflection image with the pinhole diameter to 1 AU.
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......