ZEISS
Level-adjustable Sample Holder with Insert for Piezo Stage (ELYRA)
LSM 880
18
000000-2071-464
10/2014 V_01
8
Level-adjustable Sample Holder with Insert for Piezo Stage
(ELYRA)
Caution! The adjustment screws and other part of adjustable sample holders may collide with
parts of the microscope, e.g. the condenser unit. When moving a motorized stage, avoid any
collisions. Do not install inserts while the stage is moving. Be careful when switching the
condenser front lens.
8.1
Stage Leveling Using X-Cite or Laser in Fluorescence
Fig. 5
Adjustable sample holder – schematic
view
Fig. 6
Focus control in epi-fluorescence
1.
Take a fluorescent sample like embedded beads and mount centrally over the objective. (Note: you
can also use your own fluorescent sample). Use WF illumination with the X-Cite or laser light and
focus manually onto the sample using the stand’s focus drive (Fig. 6/
A
).
2.
Set stage to zero by pressing the
Set Zero
button in the
Stage
tool. Set focus to zero by pressing
the
Manual
button in the
Focus
tool.
3.
Adjust leveling in X in the following way:
a)
Enter in the
Stage
tool in the
Step [
µ
m]
input box "1000". Press the upper or lower pointing
arrow of the
X
input box, whatever is required to step 1000 µm to the right X direction.
b)
If the sample is out of focus (Fig. 6/
B
) refocus manually by using the stand’s focus drive
C
); otherwise no further leveling is necessary. Record the relative change in the focus
position by reading the value in the
Z-Position
input box of the
Focus
tool. If the focus is off
more than 10 µm it is recommended to do a pre-adjustment. To this end adjust the edges of the
clefts viewed from the right side (Fig. 5/
csv
) and back side (Fig. 5/
cbv
) to be parallel using
screws
sy
and
sx
, respectively (Fig. 5). The cleft’s width should be approximately 2 mm as preset
by the factory, but can be adjusted with screw
sz
if needed.
c)
Multiply that value with
55
and manually set the focus onto this value using the stand’s focus
drive (Fig. 6/
D
). (Note: The factor
55
is only accurate, if the sample was positioned in the center.)
d)
Use the back left screw (Fig. 5/
sx
) of the sample holder to readjust the focus (Fig. 6/
E
). Turn
clockwise if the focus position was positive, counter clockwise if it was negative. (Note that one
turn corresponds to approximately 125 µm).
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......