Set up scan mode in the LSM
Acquisition
Mode
tool (Fig. 306). Choose
Line
scan,
highest Scan speed, 512 pixels and
Zoom 1
for
convenience.
•
Set up scan control in the LSM
Channels
tool
(Fig. 307). Open Pinhole to maximum. Set
detector gain between 200 and 300, and laser
power to 2 %.
Fig. 306
Image settings for line scan mode
Fig. 307
Channel settings for finding glass
reflection