ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
226
000000-2071-464
10/2014 V_01
(3)
PCH
In
PCH
panel the binning time used for the photo
counting histogram can be selected.
•
Click on the
PCH
tab to open the
PCH
panel
You can either check the
Automatic
check-box, in
which case 32 different binning times will be used,
or you can de-activate the check box and enter a
number of a constant binning time (in µs) in the
Binning
display window. You can either type in
values or use the arrows. Units are displayed at the
right of the display box. The default value of 10 µs
can be selected by pressing the
10
button.
In automatic binning mode, binning starts with a value of 50 ns, which is doubled 32 times. So
binning times are 50 x 2
n
, with n=1 to 32. The histogram with the best dynamic range (three
standard deviations) will be selected and displayed. In automatic binning mode, the
Binning
display window and the
10
default button will not be accessible.
(4)
Dust filter
The
Dust filter
panel allows you to activate an
electronic dust filter that will be active in operation
during the measurement. You can define a
threshold in the
Count rate
intensity that, if
exceeded, will lead to a removal of the
corresponding count rate region from the
correlation analysis (Fig. 375).
•
Click on the
Dust filter
tab to open the
Dust
filter
panel (Fig. 294).
•
Switch on or off the dust filter by ticking/
unticking the check box.
•
A threshold in % is set by typing in a value into
or by using the arrows of the
Deviation to
erase peak in Count rate
display window. All
measurement points within a binned count rate
time window having a deviation of more than
the specified value from the average count rate
will be cut out and not used for the correlation
analysis (Fig. 295). The default value of 10 %
can be activated by pressing the
10
button.
Fig. 293
Correlator tool - PCH active
Fig. 294
Correlator tool - Dust filter active
Содержание LSM 880
Страница 1: ...LSM 880 LSM 880 NLO Operating Manual October 2014 ZEN 2 black edition...
Страница 650: ......
Страница 651: ...Confocal Laser Scanning Microscopy Stefan Wilhelm Carl Zeiss Microscopy GmbH Carl Zeiss Promenade 10 07745 Jena Germany...
Страница 678: ......
Страница 687: ......
Страница 688: ......