Left Tool Area and Hardware Control Tools
LSM 880
152
000000-2071-464
10/2014 V_01
5.2.17
Multidimensional Acquisition – Tile
Scan
When the
Tile Scan
tool tab is expanded the
Tile
Scan
tool becomes available (Fig. 204).
This function permits a frame to be created as an
overview image of the specimen. The size of the
final tiled image depends on the settings for the
image format set in the
Acquisition Mode
tool
and the number of tiles.
The application of the
Tile Scan
function requires
an objective with a minimum magnification factor
of 2.5x.
The number of
Tiles
can be defined in the Tiles
Horizontal
and
Vertical
Input boxes. A maximum
of 100 tiles can be set. The size of the final Tile
Scan is displayed in
Pixels
and as a
Size
in
µ
m.
According to image format the number of
Pixels
changes. According to the objective used for
imaging the
Size
of the Tile Scan changes.
The Tile Scan grid is displayed in the
grid display
area. It shows the number and location of the
single tiles and also indicates any overlap.
The Tile scan is created around the absolute
starting position of the stage. In case the stage
does not move exactly horizontal to the scan field
a
Rotation
value can be set in the
Rotation
input
box to correct for this (available only in
Show all
mode). This can either be calibrated beforehand
using a grid slide or any other specimen which
displays clearly visible structures in horizontal
direction.
Acquire smaller tile scans to see the effect of a
specific value for
Rotation
until you find the right
one. This value should be taken for the
Rotation
.
Alternatively, use the
Tile Scan Rotation Macro
which provides an automatic
Rotation
calibration
function. Take over this value to the main software
before acquiring the tile scan.
When
Bi-directional
is active (available only in
Show all
mode) the image acquisition is
performed also on the horizontal backwards
movement of the stage.
If the system is equipped with the software option TILES, the
Overlap
input box is available (Fig. 205).
Define the overlap of the tiles as the
%
of the tile size. The extent of overlap will also be displayed in the
grid display
area. As soon as the Overlap is > 0 % the
Online stitching
checkbox becomes available.
Online stitching cannot be activated if the HDR imaging is selected in the
Acquisition Mode
tool.
Fig. 204
Tile Scan tool – Centered grid active
Fig. 205
Tile Scan tool – Overlap defined
Fig. 206
Tile Scan tool – Marked positions