ZEISS
Left Tool Area and Hardware Control Tools
LSM 880
150
000000-2071-464
10/2014 V_01
5.2.16
Multidimensional Acquisition –
Bleaching
When the
Bleaching
tool tab is expanded the
Bleaching
tool becomes available (Fig. 202).
The
Bleaching
tool is only available in LSM
Imaging mode.
The use of this function permits setting the bleach
parameters using the LSM 710 and LSM 780
systems for bleaching in spot, line or frame mode.
The
Bleach Settings
can be saved under a name
and reloaded for later use. Existing settings can
also be deleted from the list.
Bleaching works in conjunction with the
acquisition of a series of images over time. During
the sequential imaging of frames, lines or a spot,
bleaching of the specimen can be performed
according to the bleach settings defined in the
Bleaching
tool.
Start Bleaching after # scans
expects a number,
which indicates the amount of scanned frames,
lines, or spots after which then the bleaching of
the sample happens.
Repeat Bleach after # scans
expects a number,
which indicates the amount of scanned frames,
lines, or spots after which then the bleaching of
the sample is repeated.
The repeated bleaching also allows bleaching the sample to an intensity value calculated as % of the
initial intensity within the region to bleach. To achieve this indicate the percentage value in the box next
to:
Stop when Intensity drops to
.
The number of
Iterations
indicates the total amount of scans which are performed for bleaching the
selected region (see also section during each bleach process.
A
Different scan Speed
for the bleaching process can be set independently from the scan speed which
is used for imaging. A lower speed results in a longer pixel dwell time, which increases the efficiency of
bleaching.
Different Z Position
: If this option is active you can set the current stage position as the one in which
the bleaching will be done by clicking
Mark Z
. This function is only available using the
Line
or
Frame
scanning mode.
When the imaging is performed in
Spot Mode
(chosen as acquisition mode in the
Acquisition Mode
tool), a
Different XY Spot Bleach Position
can be selected. Click
Spot Select
in the
Acquisition
Mode
tool. A new image appears with a green
crosshair
. When clicking and dragging the crosshair a
red crosshair becomes visible. The red crosshair marks the spot that is imaged (This is also indicated in the
Scan Area
of the
Acquisition Mode
tool. The green crosshair marks the spot that is bleached. Move the
center of the crosshairs to the desired positions.
Use one out of four triggers with
Trigger In
to trigger the bleaching process. In addition, one out of four
triggers can be used as
Trigger Out
signal to control an external device. (For a detailed description on
the trigger use and interface see printed manual CHAPTER 4: ANNEX).
Fig. 202
Bleaching tool