Calibration for Airyscan
8. In the Airyscan box, look at the Quality
and Status. It will begin with
“Unknown (waiting)” in white, then go to
“Bad (Waiting)” or “Bad (adjusting)” in red
and when it is completed it will say
“Good (Waiting)” in Green. Once your calibration
is complete uncheck the Adjust in continuous scans.
Note: if it stays in “Unknown (waiting)” in white
for a long time your sample may not be bright
enough, or it may not be filling up enough of the
scanned area. If this happens please use one of
the slides provided by the FMIC.
25