OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-88
B 45-0021 e
03/06
Hyperfine Z Sectioning
tab
Activation of this tab is only possible if the piezo
objective focusing device has been connected.
The piezo objective focusing device can be
controlled via software (see
Stage
, page 4-122).
The accuracy of the piezo objective focusing device
regarding the step width in the Z-direction lies in
the range of 60 nm.
The piezo objective focusing device allows stacks
to be produced considerably quicker than via the
focus of the microscope stand.
The focus position remains unchanged.
•
Clicking on the
Leveling
button moves the piezo objective focus to the zero position, while the motor
focus moves into the opposite direction at the same time, i.e. the position of the object in relation to
the objective remains unchanged. This function is used to set defined initial conditions.
•
Use the slider or the arrow keys to set the number of slices for the Z Stack.
•
Use the slider or the arrow keys to set the size of the interval.
Num Slices
and
Interval
can be varied independently of each other within the piezo objective focus
work range of ±250 µm. When change is made to
Z Sectioning,
or vice versa, values are also taken
over, provided they are within the piezo objective focus work range.
If a larger range is set for the Z Stack under
Z Sectioning
or
Mark First/Last
, the
Interval
is matched
accordingly when changing to
Hyperfine Z Sectioning
, while
Num Slice
remains constant.
•
Use
XYcont
,
Line Sel
and
Range
to determine the parameters of the Z Stack (identical to Z
Sectioning).
If the green line (Current Slice) is shifted after the creation of
Range
, the focus position will change (the
piezo objective focusing device remains in the center position). The red lines (stack limits) can only be
changed symmetrically to the Current-Slice position within the piezo objective focusing device work
range.
Since the piezo objective focusing device moves from bottom to top during the creation of the
Z Stack, top and bottom of the Axiovert 200 M have been switched round.
Fig. 4-68
Hyperfine Z Sectioning tab
Содержание LSM 510
Страница 1: ...LSM 510 LSM 510 DuoScan LSM 510 META LSM 510 META DuoScan Release 4 0 March 2006...
Страница 62: ...IMPORTANT NOTES FOR CHAPTER 4 LSM 510 DuoScan Carl Zeiss LSM 510 META Duo Scan 4 II B 45 0021 e 03 06...
Страница 442: ...Programming for LSM LSM 510 DuoScan Carl Zeiss Events LSM 510 META DuoScan 5 18 B 45 0021 e 03 06...
Страница 526: ...3D FOR LSM LSM 510 DuoScan Carl Zeiss Functions LSM 510 META DuoScan 7 60 B 45 0021 e 03 06...
Страница 564: ......
Страница 566: ...MULTIPHOTON LASER SCANNING MICROSCOPY Carl Zeiss Contents LSM 510 META NLO 9 4 B 45 0021 e 03 06...
Страница 591: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 4 0 March 2006...
Страница 606: ......