Left Tool Area and Hardware Control Tools
Lightsheet Z.1
172
000000-1790-528
02/2013
3.4.5
Maintain – Adjustment
In the
Adjustment
tool window, there are three
buttons that start tools for the detection module
recognition and alignment (Fig. 269).
Detect Camera Order
,
Automatic Detector
Alignment
, and
Manual Detector Alignment
are inherently part of the detection module
exchange procedure.
When you wish to
exchange detection modules
, please follow the complete descriptions in
CHAPTER 1 HARDWARE
.
Using the show all mode for this tool window, a fourth button appears:
Initial Light Sheet Alignment
.
This is a task for authorized service personnel of Carl Zeiss Microscopy GmbH only and is therefore
password protected.
Do
not
use the
Initial Light Sheet Alignment
service task. You run the risk of altering elementary
settings for your Lightsheet Z.1 system that have been optimally set upon delivery and installation.
Please contact your local Carl Zeiss representative regarding this point only if you have designed
and built your own sample chamber.
3.4.5.1
Adjustment – Detector Recognition
When the detection modules are exchanged or attached to the Lightsheet Z.1 system for the first time,
the system must identify which detection module is present for which channel.
Pressing the button
Detector recognition
will start an identification routine to do this job. A progress
bar is visible during the process.
When the
Cancel
button is pressed, the operation is aborted without result.
This routine must be conducted before the
Automatic
or
Manual Detector Alignment
.
At the end of this process, you will be informed if the detector recognition had to perform any changes.
If so, please restart ZEN software before continuing.
Fig. 269
Adjustment tool window