Lightsheet Z.1
Carl Zeiss
02/2013
000000-1790-528
27
3.3.2
Assembly of the Sample Chamber Body and the Sample Chamber Dove Tail Slide
−
Place the mesh screen (Fig. 20/
2
) into the mid
dle of the sampl
e chamber body (Fig. 20/
1
)
to reduce
turbulence stress on the specimen when the chamber is filled.
−
For cleaning, it can be carefully removed by a forceps or via a teflon-tip-free end of a capillary
plunger.
−
If a
Heating component
is used, proceed first with chapter 3.7.1 (see PeCon manual for specific
heating component) and screw it onto the bottom of sample chamber body.
•
The sample chamber body (Fig. 21/
3
)
can now be attached to the sample chamber dove tail slide
1
).
•
Lay down the sample chamber dove tail slide with the two positioning pins facing upwards (Fig. 21/
2
).
•
Take the sample chamber body and position it on to the sample chamber dove tail slide with the
detection optic adapter window facing to the back relative to the sample chamber dove tail slide grip.
•
Invert the sample chamber body with the sample chamber dove tail slide so that the four screw-holes
(Fig. 21/
4
)
that were on the bottom are now facing you.
•
Insert the four screws (diameter = M3, length = 14 mm) (Fig. 21/
5
)
into the four screw-holes and
tighten using and Allen wrench (2.5).
1
Mesh screen
2
Sample chamber body
Fig. 20
Placing the mesh screen