Set focus for Cam 1 (Channel 1)
•
Use the
Light Path
of the
Manual Detector Alignment
to first activate channel 1 (Cam1) and press
afterwards the
Continuous
button. The resulting live image shows the grating on channel 1 (Cam1).
•
Use the
Zoom
tools of the Manual Detector Alignment to zoom into the center region of the image.
•
With the live image running, open the
front system door
1
) to the
reflector turret for
emission filter selection
. This system door has no safety interlock, and opening it will not terminate
image acquisition.
•
Turn the screw which is above the turret (Fig. 14/
3
) with the long Allen wrench (Fig. 14/
2
).
•
Monitor the live image of the grating while you turn the screw in any direction and find the position
of the screw that results in the sharpest image.
•
Stop the continuous scan and close the
front system door
.
Set focus for Cam 2 (Channel 2)
•
Use the
Light Path
of the
Manual Detector Alignment
to direct light towards channel 2 and
activate channel 2 (Cam2).
•
Press afterwards the
Continuous
button. The resulting live image shows the grating on channel 2
(Cam2).
•
Use the
Zoom
tools for the Manual Detector Alignment to zoom into the center region of the image.
•
While the imaging continues remove the black screw cap (Fig. 15/
1
) on the back of the main system
module Lightsheet Z.1. This opening gives way to a screw that can be turned with an Allen wrench
(Fig. 15/
2
).
1
Front system door
2
Allen wrench
3
Screw