ZEISS
2 Requirements | 2.3 Mains Connection
2.3 Mains Connection
The Microscope System must be properly installed with the power cords and supplied socket with
protective ground contact securely connected.
The provided electrical connection must be in accordance with the applicable electrical codes for
the country of installation. In order to avoid disturbance from other installed machines, you must
use a separate power connection.
§
Make sure the mains connector is not more than 1.5 m away from the position of the laser
rack.
§
Do not extend or modify the supplied 2.5 m power cord.
§
Do not connect electrical systems that are not authorised by ZEISS on the supplied power sup-
ply cord.
Value
Nominal AC voltage
1/N/PE 220-240 V AC (+/- 10%) / 100 -
125 V AC (+/- 10%)
Nominal frequency
50 - 60 Hz
Main Power Plug
Local mains plug will be supplied
Power consumption
(220-240 V) 800 VA
(100-125 V) 750 VA
Storage PC: Max. 400 VA
Max. current
3,5 A at 220 V
8 A at 100 V
Additional building PE
The system must be connected to a building
earth point at all times.
Additional mains power
3 x Standard local 230 or 120 V mains sock-
ets
Required for the Storage PC, Monitor and
other accessories.
Circuit breaker type
Type C
MCB (Miniature Circuit Breaker) or
MCCB (Moulded Case Circuit Breaker) ac-
cording to IEC EN 60898
Leakage current (relevant for Residual Cur-
rent Device)
Max. 23 mA at 230 V
Max. 11 mA at 120 V
Protection class
IP20 (IEC 60529)
IEC earth class
Class 1 of IEC 61140
All chassis are connected to electrical earth
by the earth cable in the mains cable.
II
Installation Requirements ZEISS Lightsheet 7 | en-US | Rev. 1 | 349500-8186-000
9