Rev. D 08/21/06 PN 701692
Humphrey
®
Matrix Quick Reference Guide
From the
Main Menu Screen (F1),
select one of the
Test
Type
Buttons or the
Select Patient
Button with the mouse
pad or Hot Key (Alt+ underlined letter). Both selections
proceed to the
View Patients (F2) Screen
.
From the
View Patient Screen (F2)
, select an existing
patient from the patient listing displayed on the screen by
clicking once on the patient name. Clicking the
Run Test
Button brings you to the
Testing Screen
.
or
You may add a new patient to the testing database by
selecting the
Add New Patient
Button in the
View Patient
(F2) Screen
. Required fields (
DOB
in addition to
First
and
Last Name
or
ID
) are marked with an asterisk (*). Once the
new patient data has been added to the testing database,
the Testing Screen will be displayed.
In the
Testing Screen
, check to be sure the proper patient
name and test type are displayed. The test type can be
changed in the testing screen if necessary. The status box in
the lower corner of the screen will state that the
Pre-test
Demo
is being performed so the patient is able to view a
demonstration of selected test.
•
Place the
Patient Visor
so that the Patient Eyepiece is
aligned with the eye to be tested.
•
Adjust the position of the patient or the instrument
to obtain proper, comfortable patient alignment
.
Proper alignment requires the patient to see all four
self-alignment points at the same time while fixating on
the black square target in the center. It is OK if the
fixation target appears fuzzy to the patient.
•
When
Eye Monitoring
is turned on, an image of the
patient’s eye will be displayed on the Testing Screen.
The
patient’s pupil should be kept inside the circle
on
the video image throughout the test.
The pupil does
not have to be perfectly centered
for proper alignment,
but it should stay within the circle.
•
After explaining the test to the patient, click the
Start
Test
Button to begin the test. During testing the Pause
Test Button can be used to pause the test. The Cancel
Test Button provides the option to cancel the test.
Select Test Type
Select Existing Patient or Add New Patient
Start Test
Keep patient’s pupil INSIDE the circle.
Содержание Humphrey matrix
Страница 25: ...Humphrey Matrix Visual Field Instrument 21 Rev D 08 21 06 PN 701692 Figure C...
Страница 63: ...Humphrey Matrix Visual Field Instrument 59 Rev D 08 21 06 PN 701692...
Страница 64: ...60 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Страница 65: ...Humphrey Matrix Visual Field Instrument 61 Rev D 08 21 06 PN 701692...
Страница 66: ...62 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Страница 67: ...Humphrey Matrix Visual Field Instrument 63 Rev D 08 21 06 PN 701692...
Страница 68: ...64 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Страница 69: ...Humphrey Matrix Visual Field Instrument 65 Rev D 08 21 06 PN 701692...
Страница 70: ...66 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...
Страница 71: ...Humphrey Matrix Visual Field Instrument 67 Rev D 08 21 06 PN 701692...
Страница 72: ...68 Humphrey Matrix Visual Field Instrument Rev D 08 21 06 PN 701692...