Start-up
Illuminator
Carl Zeiss
Inserting and replacing the burner
HBO 50
12
M60-2-0014 e 05/2012
•
Press spring lever (3/
6
) of cooling element
up and pull cooling element of transport
safety device or of used burner (3/
3
) up and
put it cautiously on the table.
•
Press spring lever (3/
5
) for lamp socket (3/
4
)
down and pull transport safety device or
burner (3/
3
) upwards out of lamp socket.
Let go of the spring lever.
•
Press spring lever (3/
5
) for lamp socket (3/
4
)
down once more and insert burner into
lamp socket. Do not touch glass bulb of
burner. Let go of spring lever.
•
Press spring lever (3/
6)
for cooling element
up and put cooling element on burner.
Caution:
You should absolutely observe
that the cooling element rests completely on
the burner until the limit stop.
•
Insert lamp compartment into lamp housing
and pull it until it clicks into place. Tighten
lock screw (3/
7
).
•
Read and note the result of the hour meter
(power supply). After you reached the
nominal life time of 100 hours of burner,
replace it.