ZEISS
9 Technical Data and Conformity | 9.5 Installation Requirements
9.5.1 Layout and Connections
1
2
3
4
A
J
I
B
E
C
D
F
G
H
1
Static vibration damper
A
Mains power supply
208...230 V/25 A, L1/N(L2)/PE
2
Pre-vacuum pump
Equipotential bonding bar
3
Computer workplace
Pressure reducer (water, nitrogen,
compressed air)
4
Emergency Off (EMO) button (optional)
Main shut-off valves
Water supply
Water runback
Nitrogen supply
Compressed air supply
Exhaust line
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000