Operation
Fluar Illuminator Z mot. Switching on and adjusting the Fluar Illuminator Z mot.
Carl Zeiss
05/2012
435180-9060-701
19
Power switch
5
):
−
Switches the HXP 200 C on and off
−
Integral LED lights up when switched on
For further instructions for the HXP 200 C refer to the manufacturer's instructions
4.3
Switching on and adjusting the
Fluar Illuminator Z mot.
The Fluar Illuminator Z mot. is switched on
together with the Axio Zoom.V16 microscope
when it is connected to the focusing motor or
EMS 3 controller by CAN-RJ45 cable.
If the Fluar Illuminator Z mot. is
supplied by its own separate plug-in
power unit, it is automatically switched
on as soon as the latter is connected to
the main supply.
•
Switch the microscope on by pressing on the
HIP.
After switching on, the shutter of the Fluar
Illuminator Z mot. is always closed.
•
Completely open the iris diaphragm of the stop
slider (man. or mot.) using the control knob
(Fig. 10/
3
) or setting wheel (Fig. 10/
6
) as
required. Check that the diaphragm is centered.
If necessary, these can be centered with the two
adjusting screws (Fig. 10/
1
and
4
) or (Fig. 10/
5
and
7
).
•
Adjust the microscope to the specimen initially without fluorescence excitation according to the
separate operating manual. For this purpose swivel into a position on the reflector turret without
fluorescence filter set.
•
Switch the HXP 200 C illuminator on by the
Power
switch (Fig. 9/
5
) and select the desired
Intensity
1
) on the control dial.
•
Set
Shutter open/remote
2
) to the
open
position.
The
Shutter open
display (Fig. 9/
4
) on the HXP 200 C lights up.
Fig. 10 Manual (right) and motorized (left)