Technical specification
Instructions for Use
11.8.1
Electromagnetic emission
The EXTARO 300 is intended for operation in an electromagnetic environment as specified below. The
customer or the user of EXTARO 300 is responsible for ensuring that the device is operated in such an
environment.
Emission test
Compliance
Electromagnetic environment- guidelines
RF emissions as per CISPR11 Group 1
The EXTARO 300 uses RF energy only for its
internal functions. As a result, RF emissions are
very low and unlikely to cause any interference
in nearby electronic devices.
RF emissions as per CISPR11 Class A
The EXTARO 300 is suitable for use in
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions as per
IEC 61000-3-2
Class A
The EXTARO 300 is suitable for use in all
facilities including locations in residential
environments and those directly connected to
the public power supply network which also
supplies buildings used for residential
purposes.
Emission of voltage
fluctuation /flicker as per IEC
61000- 3-3
Compliant
11.8.2
Electromagnetic immunity for ME equipment and ME systems
The EXTARO 300 is intended for operation in an electromagnetic environment as specified below. The
customer or the user of the EXTARO 300 is responsible for ensuring that the device is operated in such
an environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic
environment – guidelines
Electrostatic dis-
charge (ESD) as per
IEC 61000-4-2
±8 kV
contact discharge
±15 kV
air discharge
±8 kV
contact discharge
±15 kV
air discharge
Floor stand should be made of
wood or concrete or be
covered with ceramic tiles. If
the flooring contains synthetic
materials, the relative humidity
must be at least 30%.
Fast transient/burst
immunity as per IEC
61000-4-4
±2 kV for
power supply lines
±1 kV for
input/output lines
±2 kV for
power supply lines
±1 kV for
input/output lines
The quality of the supply
voltage should be that of a
typical business or hospital
environment.
Surges as per IEC
61000-4-5
± 1 kV
phase-to-neutral
voltage
± 2 kV
phase/neutral to
ground voltage
± 1 kV
phase-to-neutral
voltage
± 2 kV
phase/neutral to
ground voltage
The quality of the supply
voltage should be that of a
typical business or hospital
environment.
51 / 59
305935-3500-000 – 2.1 – 2017-12-15