5 First Operating Steps | 5.8 Switching Off the Microscope System
ZEISS
5.8 Switching Off the Microscope System
5.8.1 Finishing the Work Session | LaB₆ Filament
To obtain the maximum lifetime of a LaB₆ emitter, switch off the EHT when the microscope is not
in use and switch the gun to
STANDBY
mode when the microscope is not used for longer peri-
ods of time. In
STANDBY
mode, the filament continues to be heated, and the vacuum in the
electron column and in the specimen chamber are maintained. It is recommended that you do not
turn off the gun during its service lifetime.
The
STANDBY
mode is also the recommended mode to store the microscope. In this case, acti-
vate the
Partial Vent on Standby
checkbox in the
Vacuum
tab of the
SEM Controls
panel. This
can help to prevent oil vapors from penetrating into the specimen chamber during the period of
storage.
To further extend the service life time, you can activate the
Long Fil. Life
checkbox in the
Gun
tab of the
SEM Controls
panel. This reduces the filament current to “first peak” conditions,
which provides satisfactory operating conditions for general microscopy. If optimum performance
is required, the
Long Fil. Life
checkbox should be deactivated.
Prerequisite
ü
The microscope is in ON mode.
Procedure
1. Click the
All:
button in the status bar.
2. Select
EHT Off
from the pop-up menu.
3. Only when interrupting work for longer periods between 2 and 7 days: In the
Vacuum
tab
activate the
Partial Vent on Standby
checkbox.
à
This maintains the gun vacuum, and switches off and protects the turbo pump.
4. Exit the SmartSEM user interface. Refer to
Closing the SmartSEM User Interface [
5. Close all programs and software.
6. Select
Windows start button > Power icon > Shut down
.
7. At the front of the plinth, press the
Standby
button.
à
The microscope switches to Standby mode.
5.8.2 Finishing the Work Session | Tungsten Filament
Info
If you operate the microscope with a tungsten filament, then the EHT and the gun are always
switched on or switched off together. You cannot separately switch on or off the EHT or the
gun.
Info
The tungsten filaments have only a limited lifetime of 100–300 operating hours. If you operate
the microscope with a tungsten filament, then do the following:
4
If the microscope is not in operation, then always switch off the EHT and gun to preserve
the filament.
Prerequisite
ü
The microscope is in ON mode.
108
Instruction Manual ZEISS EVO | en-US | Rev. 10 | 354706-0780-006