CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
DirectFRAP
Mounting the Excitation Filter Wheel 8x mot.
Carl Zeiss
05/2014
423635-9001-701
57
5.9
Mounting the Excitation Filter
Wheel 8x mot.
•
Remove the cover of the reflected light mount
(Fig. 5-11/
1
) of the microscope.
•
Insert the filter wheel with the dovetail mount
(Fig. 5-11/
6
) into the reflected light socket
1
) and tighten the locking screw
2
) using the 3 mm ball-end Allen key.
•
Mount the sensor ring of the safety kit for
HXP 120 (Fig. 5-11/
7
, right) at the mount
3
) of the filter wheel.
•
Insert the plug (Fig. 5-11/
4
) of the filter wheel
into the CAN port of the microscope.
•
Insert the optical fiber connector of the
HXP 120 V with mounted contact ring
(Fig. 5-11/
7
, left) into the mount (Fig. 5-11/
3
) of
the filter wheel and fix it with the clamping
screw (Fig. 5-11/
5
).
5.10
Changing the HAL Halogen Lamp
For changing the halogen lamp of the HAL illuminator, the laser must be switched off in any
case!
See separate operating manual.
Fig. 5-11
Mounting the filter wheel