7 Operation | ​

ZEISS

7 Operation

 WARNING

Burning hazard

Danger of burns or frostbites when liquid nitrogen gets in contact with the skin.

4

Wear proper protection equipment when handling the liquid nitrogen and Cryo Trap.

4

Do not touch the cooling rod and the cooling plate inside the chamber as long as the de-
war is not empty.

4

The dewar has to be in upright position. Never unscrew the dewar as long as liquid nitro-
gen is in it.

4

Use suitable tools when filling up the dewar.

4

The dewar must be empty and must have room temperature before exchanging the cool-
ing plate of the Cryo Trap.

 WARNING

Explosion hazard

Explosion risk if the pressure inside the dewar rises due to evaporating liquid nitrogen.

4

Only use the cover which is provided by ZEISS for the dewar filling hole.

4

Regularly check that ice does not build up at or around the cover for the dewar filling hole.

 WARNING

Suffocation hazard due to lack of oxygen

Danger of nitrogen accumulation in low-lying areas or subterranean rooms.

4

Ensure the room is ventilated properly.

4

Ensure the room is equipped with a ventilation system with bottom suction and air ex-
changer.

 WARNING

Fire hazard

Fire risk due to concentration of oxygen in liquid nitrogen.

4

Cover the dewar so that the evaporating nitrogen can escape, but the penetration of air is
inhibited and no flammable material can fall into the dewar.

4

Store and transport liquid nitrogen only in suitable dewars.

4

Do not smoke or have open fire in the same room the dewar is located.

NOTICE

The detectors can be damaged

The cooling plate may collide with the detectors during mounting because there is very little
space between cooling plate and detectors.

4

The EDS detector should be retracted during installation.

4

Take care of the nearby detectors during the whole mounting procedure.

Info

14

Instruction Manual ZEISS Cryo Trap | en-US | Rev. 2 | 346000-8166-000

Содержание Cryo Trap

Страница 1: ...Instruction Manual ZEISS Cryo Trap...

Страница 2: ...for archiving purposes shall remain unaffected thereby Any viola tions may be prosecuted as copyright infringements The use of general descriptive names registered names trademarks etc in this docume...

Страница 3: ...p 9 5 2 Principle of Operation 10 5 3 Technical Data 11 5 4 System Requirements 11 5 5 Detector Compatibility 12 5 6 Stage Movement Limitiation 12 6 Installation 13 7 Operation 14 7 1 Filling the Dewa...

Страница 4: ...al before starting to work with the Cryo Trap Read the instructions carefully Keep the instruction manual nearby the Cryo Trap and hand it over to future owners of the instrument This instruction manu...

Страница 5: ...Follow a path in the software Text input Text output Text to be entered by the user Text displayed by the system Programming and Macros Anything typed in literally during program ming including for e...

Страница 6: ...rod connecting the cooling plate to the dewar Contamination Unwanted absorbation of volatile hydrocarbon compounds at speci men and instrument surfaces Cryo Trap Reduces the contamination in the spec...

Страница 7: ...n contact with the skin 4 Wear proper protection equipment when handling the liquid nitrogen and Cryo Trap 4 Do not touch the cooling rod and the cooling plate inside the chamber as long as the de war...

Страница 8: ...long as the dewar is filled with liquid nitrogen 4 Before you vent the specimen chamber wait until the dewar is empty and the cooling rod and cooling plate are warmed up to room temperature 4 Always...

Страница 9: ...r 4 Adapter Flange 5 Cooling Plate 6 Cooling Rod Function The chamber vacuum provides the thermal isolation of a dewar vessel which is filled with liquid nitrogen The extreme cold will be transferred...

Страница 10: ...comes contaminated This may influence the image quality 1 2 The temperature of the cooling plate 3 decreases and the hydrocarbons change their physical state from volatile to solid The compounds are d...

Страница 11: ...he cooling plate stays below 130 C for 10 12 hours Warming up time At regular system vacuum With a refilled dewar it takes about 2 days to reach room temperature again 36 hours to reach 20 C CryoWarmU...

Страница 12: ...NT The EDS detector must be mounted on port EDS1 Fig 2 Ports of GeminiSEM 300 500 5 6 Stage Movement Limitiation With cooling plate in position untilted stage movement is uncompromised A stage tilt of...

Страница 13: ...ZEISS 6 Installation 6 Installation The Cryo Trap is either factory installed or retrofitted by authorized ZEISS service personnel Instruction Manual ZEISS Cryo Trap en US Rev 2 346000 8166 000 13...

Страница 14: ...he cover for the dewar filling hole WARNING Suffocation hazard due to lack of oxygen Danger of nitrogen accumulation in low lying areas or subterranean rooms 4 Ensure the room is ventilated properly 4...

Страница 15: ...e lid white Teflon cover 6 Make sure that the lid is correctly seated and the borehole is not facing towards the speci men chamber 7 For extended operating hours 10 hours refill liquid nitrogen as nec...

Страница 16: ...objective lens 3 Fix the cooling plate with the knurled thumb screw 2 2 4 Measure the resistance between SE2 collector grid and the cooling plate with a DVM These two parts should be isolated from ea...

Страница 17: ...installation and Disposal 8 De installation and Disposal The Cryo Trap is de installed and disposed of by authorized ZEISS service personnel only Instruction Manual ZEISS Cryo Trap en US Rev 2 346000...

Страница 18: ...om temperature Do not attempt to pump the cham ber again until all components are warm and fully dry Defrost the dewar Image drift Fluctuating temperature in the specimen chamber Check position of bor...

Страница 19: ......

Страница 20: ...Carl Zeiss Promenade 10 07745 Jena Germany phone 49 3641 64 3161 fax 49 3641 64 3439 info microscopy zeiss com www zeiss com microscopy Instruction Manual ZEISS Cryo Trap en US Rev 2 Printed in Germa...

Отзывы: