Lamp replacement and calibration - Corona process | Service manual |
EN_40_070_0010I
Sensor calibration
Lamp replacement and calibration Corona process
5 Sensor calibration
•
at least
InProcess V 2.7.0
installed on PC.
Download the latest version under “
https://download.zeiss.com/
spectroscopy/software/InProcess/”
.
•
both halogen lamps replaced and the device fully reassembled
•
Tools and accessories:
– Corona process calibration assembly, order no. 000000-2211-998
Delivery package: � Calibration device
� certified white standard with corresponding certificate
(accompanying CD)
� Black standard
� Tools for the lamp change
� Associated instructions
� Storage box
– Ethernet cable, order no. 000000-2036-573
– Lab power supply, order no. 000000-2045-548
1
Assemble calibration device
Open the calibration device.
Prepare the reference standards.
2
Prepare measuring system for the calibration
Mount the
Corona process
on the calibration device.
Fig. 29 Mounting the Corona process on the calibration device
Requirements
Procedure