Replacing halogen lamps
Lamp replacement and calibration Corona process
4.5 Installing new halogen lamps
Caution
RISK TO FUNCTIONALITY
For devices with hardware revision 1 the lamp replacement must be
done by Carl Zeiss Spectroscopy GmbH. Device must be sent to the
service department.
Hardware revision 2 requires installation of lamp type B.
Caution
SENSITIVE COMPONENTS
When mounting a new lamp
4
, ensure that
the lamp cable
6
is not
crushed, bent, or otherwise damaged.
Note
All action steps described in this paragraph refer to lamp type B.
1
Install new halogen lamp
Put on the cotton gloves.
Carefully guide the lamp cable
6
of the new halogen lamp
4
through the
lamp holder
5
.
Fig. 26
4
5
6
Guiding lamp cables through lamp holder
Use the new halogen lamp
4
in the lamp holder
5
.
Procedure