27
EN_40_070_0010I
|
Service manual | Lamp replacement and calibration - Corona process
Replacing halogen lamps
Lamp replacement and calibration Corona process
Note
Some cables pass through the lamp holder of the L2 spare lamp.
Therefore the lamp holder must be removed laterally.
4
Removing the halogen lamp
Place the lamp holder
5
on a clean surface.
Remove the three screws
1
from the lamp holder
2
using a
TX 10 screwdriver and remove the clamping ring
3
.
Remove the old halogen lamp
4
from the lamp holder.
Fig. 25
1
3
2
4
4.1
6
5
Removing the halogen lamp – lamp type B