Probing the workpiece
7-20
61211-1310402
CONTURA
Probing the workpiece
How to ensure correct measurement
NOTE
Regular inspection of the stylus systems is required to ensure correct
workpiece measurement.
–
If particles adhere to the stylus tip or a lubricating film has formed,
the stylus tip must be cleaned.
–
When the stylus has become damaged, it must be replaced.
–
The adapter plate must likewise be free of impurities and must be
cleaned, if necessary.
What you should know!
The type of probing depends on the measuring job and the required
number of measuring points. Position measurement requires only one or
more discrete-point probing actions. Multipoint probing or scanning can
be used when determining dimensions. Scanning must be used for form
measurements.
NOTE
The following conditions must be fulfilled for probing: These conditions
depend on the probing system. See brochure »Sensors«.
Probing conditions
–
Constant probing speed.
–
Minimum distance to the contact point.
A minimum distance is required to reach the probing speed. The
minimum distance should be 2 to 4 mm. In case of smaller dis-
tances, the probing time increases.
Probing speed
Probing should be carried out at constant travel speed.
The travel speed depends on the deflection of the joystick. At the begin-
ning and end of the deflection range, the travel speed is proportional to
the amount of deflection. The travel speed between the beginning and
the end is constant. This speed corresponds to the probing speed.
Содержание CONTURA
Страница 46: ...Components and function 3 10 61211 1310402 CONTURA Attaching the holders to the profile rail 6 9...
Страница 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Страница 62: ...CMM 4 8 61211 1310402 CONTURA...
Страница 68: ...Installation 5 6 61211 1310402 CONTURA...
Страница 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Страница 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Страница 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Страница 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Страница 130: ...Service features 8 8 61211 1310402 CONTURA...
Страница 131: ...9 61211 1310402 CONTURA 9 1 Maintenance and care This chapter contains Maintenance 9 2 Care 9 4 Maintenance and care...
Страница 150: ...2Glossary 2 61211 1310402 CONTURA...
Страница 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Страница 157: ......
Страница 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...