START-UP
Cell Observer SD
Installation Requirements
Carl Zeiss
M60-2-0021 e 01/2009
423638-7044-001
37
3
START-UP
The Cell Observer SD is a laser class 3B system. It is provided with respective labeling (total
maximum output power of 300 mW). You are obliged to meet the provisions of health
protection and workplace safety when handling laser devices. In addition, users must comply
with all applicable provisions of statutory law.
Notify your Laser Safety Officer before proceeding with the initial Cell Observer SD system start-
up.
The Cell Observer SD system may not be installed, retrofitted or adjusted by anyone other than
service personnel of Carl Zeiss MicroImaging GmbH or persons duly authorized to do so by the
Manufacturer. Intended and conforming use provided, the Cell Observer SD can be
commissioned and operated by Customer personnel without further safety precautions (refer to
section 2.1 on page 20).
Before installing and starting up the instrument it is imperative to read carefully the Fehler!
Verweisquelle konnte nicht gefunden werden..
3.1
Installation Requirements
3.1.1
Space Requirements
Fig. 3-1 shows the floor space required for system installation:
800
Fig. 3-1
Space requirements
For system installation, the following requirements must be met:
−
floor area required: 7 m
2
,
−
stable laboratory worktable with system on vibration-isolated base plate or vibration-isolated, air-
cushioned microscope table (with dimensions of 180 x 90 cm),
−
space between system and wall: min. 50 cm.