START-UP
Axioskop
2
plus
Carl Zeiss
Initial start-up
Axioskop 2
mot plus
2-6
B 40-075 e 10/03
2.1.4
Setting of the interpupillary
distance of the binocular tube
The eyepiece distance is matched to the user's
individual interpupillary distance by swinging the
eyepiece tubes symmetrically towards one another
(2-5/
A
).
2.1.5
Setting of viewing height
The viewing height can be set continuously in the
range from 6 to 25° by changing the viewing
angle of the two ergotubes. Furthermore, the
binocular component of the ergotubes can be
swivelled by 180°.
2.1.6
Screw in of objectives
•
Move mechanical stage with carrier to the
lower stop position.
•
2
) from the respective
openings in the nosepiece.
•
) from case and screw
them in nosepiece (2-6/
) clockwise one by one,
starting with the lowest magnification. In the
case of the Axioskop 2
mot plus
, start with
objective position No. 1 of the coded nosepiece.
☞
Make sure to use the dust
protection caps on those nosepiece
eyes which are not currently
required.
Fig. 2-5
Setting the interpupillary distance
of the binocular tube
Fig. 2-6
Screwing in of objectives