![Zeiss Axio Vert.A1 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/zeiss/axio-vert-a1/axio-vert-a1_operating-manual_927645054.webp)
PUTTING THE INSTRUMENT INTO OPERATION
Carl Zeiss
Mounting Standard Components
Axio Vert.A1
54
431030-7044-001
05/2012
3.1.11.2
Adjusting the HBO 50 Illuminator
WARNING
To prevent any irreparable eye
damage, never look directly into the
ignited lamp. Put on protective
glasses, such as sunglasses, to protect
your eyes when observing the bright
light spot.
•
Insert a filter set for GFP (green fluorescent
protein).
•
Switch on the fully connected illuminator on the
power supply unit. The burner will ignite
automatically.
•
Screw out an objective. Using a free passage,
check the source image on a sheet of paper
held at a distance of approximately 10 mm
above the objective shoulder.
•
Using the adjusting screw (Fig. 3-17/
1
), set the
burner in axial direction to the mirror until the
two light arcs appear equally sized on the
source image (see Fig. 3-17/
4
).
•
Position the light arc and the mirror image
parallel to one another (Fig. 3-17/
4
) with the aid
of the height (Fig. 3-17/
3
) and lateral
adjustment (Fig. 3-17/
2
) screws, respectively.
The light arc and the mirror image should
overlap by one third.
Fig. 3-17
Adjusting HBO 50