3.1.11.1
Attaching the Fluorescence Illuminator for Reflected Light
For safety reasons, the HBO 50/HBO 100 fluorescence lamps and the HXP 120C mercury vapor short-arc
lamps are packaged and delivered to the customer separately.
Before those lamps can be put into operation, they must, therefore, be inserted into the illuminator
housings.
The procedure to follow when installing or replacing the lamp is described in the operating instructions
enclosed with the devices.
To install the illuminators, proceed as follows:
•
Loosen the clamping screw (Fig. 3-16/
3
) and remove the protective cap from the reflected light port
2
).
•
Insert the HBO 50 (Fig. 3-16/
1
) or the adapter for the HXP 120C illuminator with the dovetail directly
into the reflected light port (Fig. 3-16/
2
) and lock it into position with the clamping screw (Fig. 3-16/
3
),
using ball-headed screwdriver SW 3.
•
When it comes to the HBO 100, be sure to attach the 35-mm distance ring (Fig. 3-16/
4
) onto the
reflected light port in the same manner before mounting the illuminator (Fig. 3-16/
5
).
•
Connect the multipole illuminator plug (Fig. 3-16/
6
or
7
) to the socket of the appropriate power
supply unit and secure it with the retaining ring.
•
First connect the power supply cable to the HBO 50/HBO 100 power supply unit and then to the
power socket.
•
Use the ON/Off switch (Fig. 3-16/
9
to turn on the power supply unit for the HBO 50 or the ON/Off
8
) to activate the HBO 100.