START-UP
Axio Imager
Activating/deactivating the high-performance focus
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
49
The reflective (coated) side (3-27/
3
) of
the beam splitter has a beveled edge
(3-27/
1
2
).
•
Place the
excitation
half of the module
1
) onto the
emission
half (3-26/
4
)
(retaining pins 3-26/
5b
and eyelets 3-26/
5a
mesh with one another). Hold both halves
together and turn them back into the
installation position.
•
Re-insert the slotted screws and tighten them
up.
•
Finally, affix the adhesive label with the name of
the filter combination to the side of the module.
3.22
Activating/deactivating the high-
performance focus
When a scanning stage (approx. weight: 4 kg) is
used, the high-performance focus must be
activated to protect the focusing drive.
Thus, specimens weighing up to 5 kg may be
placed on the stage for microscoping.
To activate/deactivate the high-performance focus,
proceed as follows:
•
Use the Z drive to move the microscope stage
into the lowest position possible.
•
Use ball-headed screwdriver SW 3 (3-28/
2
) to
screw the two screws (3-28/
1
), located to the
right
and
left on the stand, in or out until they
reach their respective end positions.
Screws screwed out: High-performance focus is
activated.
Screws screwed in:
High-performance focus is
deactivated.
Fig. 3-27
Marks on beam splitter
Fig. 3-28
Activating/deactivating the high-
performance focus