START-UP
Axio Imager
Equipping 2-position filter wheels 2x, discrete
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
45
3.18.2
Filter wheel, 2-position, motorized
The motorized filter wheel 2x is to be equipped
with the same filter set as the manual filter wheel
2x.
To insert the filters in the two filter wheels, the
motorized filter wheel 2x must be opened on the
corresponding side.
The filter mounts are labeled with the
corresponding position numbers 1 to 4 (3-21/
1a
to
4a
and
1b
to
4b
).
•
Loosen both screws (3-21/
6
and
7
) and remove
cover plate (3-21/
5
and
6
).
•
Take care that no dirt gets onto the filter
surface.
•
Screw in the retaining ring.
•
After all filters have been inserted, put on the
cover plate and screw it down.
In filter wheel 1 (428301-9901-000; Fig. 3-21),
insert the following filters:
Position
1a
:
No filter (100% transmission)
Position
2a
:
Neutral-density filter 12%
Position
3a
:
Neutral-density filter 25%
Position
4a
:
Neutral-density filter 50%,
reflective coating up
In filter wheel 2, insert the following filters:
Position
1b
:
No filter (100% transmission)
Position
2b
:
Neutral-density filter 6%
Position
3b
:
No filter (100% transmission)
Position
4b
:
Neutral-density filter 0.5%
and
neutral-density
filter
25%,
reflective
coating
down
Fig. 3-21
Equipping filter wheel 2x,
motorized (only for
428301-9901-000)