START-UP
Carl Zeiss
Setting the dovetail guide of the stage carrier
Axio Imager
42
430000-7344-001
M70-2-0020 e 06/2009
3.15
Setting the dovetail guide of the
stage carrier
When stage carrier 430701-9901-000 is used, the
specimen space needs to be shifted for specimens
of less than 13 mm, so that the specimen can be
focused.
To do so, proceed as follows:
•
Remove the stage carrier (see Section 3.13).
•
Loosen
the four locking screws (
1
to
4
)
of the dovetail guide.
•
Move the guide plate (3-17/
5
) as much
upward
in the slots as it will go.
•
Then, tighten the four screws.
Always
proceed
in the following order:
−
upper left screw (
1
),
−
lower left screw (
2
)
,
−
upper right screw (
3
)
,
−
lower right screw (
4
)
,
•
Attach the stage carrier (see Section
If stage carrier 430701-9901-000 is no longer used, the guide plate must be set back to the
lowe
r position (normal position).
Fig. 3-17
Setting the dovetail guide up for
specimen space expansion