FIRST-TIME SET-UP
Axio Imager 2
Installing the focus linear sensor
56
430000-7544-001
01/2016
3.26
Installing the focus linear sensor
Only in connection with
Axio Imager.Z2 / .Z2m stands.
Adjusting block 430702-0102-000 is
required for installation.
•
Switch on the microscope (see Section 4.3.1).
•
Screw the objective which is in the optical path
out of the nosepiece.
•
Move the mechanical stage with stage carrier to
the upper stop position.
•
Apply the adjusting block (Fig. 40/
6
) to the
stage carrier and fix the top part of holder Z
(Fig. 40/
3
) on it.
•
Bring the top part of holder Z (Fig. 40/
3
) into
line with the adjusting block and fasten it to the
stand using both screws (Fig. 40/
2
).
•
Press the bottom part of holder Z (Fig. 40/
5
)
against the adjusting block, so that the stage
carrier and the bottom part of holder Z are
flush.
•
Fasten the bottom part of holder Z (Fig. 40/
5
) with both screws (Fig. 40/
4
) to the stage carrier.
•
Insert the focus linear sensor (Fig. 40/
1
) in the top part, bring it into line with the lower edge and
fasten it by means of clamping screw (Fig. 40/
7
).
•
Move the stage carrier to the lower stop position and remove the adjusting block.
•
Switch off the microscope (see Section 4.3.2).
•
Plug the cable of the focus linear sensor into connector (Fig. 42/
7
) on the rear side of the stand and
switch on the microscope (see Section 4.3.1).
Incorrect installation of the focus linear sensor can cause problems in focusing the object.
The focus linear sensor can be activated and deactivated via the TFT display (see Section
on page 133).