CARE, MAINTENANCE, TROUBLESHOOTING AND SERVICE
Axio Examiner
Equipment maintenance
Carl Zeiss
M60-2-0003 e 05/2012
83
•
Shut down and secure the equipment whenever a damage is found. Have damage repaired by
qualified specialist personnel.
•
Check that the maximum operating time is observed for halogen or mercury vapor short arc lamps.
5.2.2
Replacing fuses
Make certain to pull out the power plug
before replacing any fuse.
The fuse box is located at the rear side of the
corresponding power supply or examiner control
unit. It is combined with the appliance inlet, and
contains two fuses of the following types:
T 5.0 A/H 5x20mm
(Examiner control and 12 V DC 100 W external
power supply unit)
or
T 2.0/H A
(HBO 100 transformer).
or
T 4 A
(mbq52ac-z 100 V, 127 V transformer)
T 2.5 A
(mbq52ac-z 220 - 240 V transformer)
•
Pull out the power plug.
•
Pull out the fuse holder (Figure 5-1/
2
) towards
the front side, and take it completely out of the
fuse box (Figure 5-1/
1
).
•
Pull defective fuses out of the fuse holder, and
replace them by new fuses. Make certain that
the fuse has the proper fuse designation.
•
Insert the fuse holder into the fuse box again,
and push it in up to the stop.
•
Plug in the power plug.
Figure 5-1 Replacing fuses, e.g., at the 12 V
DC 100 W external power supply
unit
Figure 5-2 Replacing fuses at the examiner
control
Содержание Axio Examiner
Страница 1: ...Operating Manual Axio Examiner Research Microscope...
Страница 13: ...INTRODUCTION Axio Examiner List of interfaces Carl Zeiss M60 2 0003 e 05 2012 13 Figure 1 2 List of interfaces...
Страница 19: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 19 2 2 System overview...
Страница 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Страница 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Страница 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Страница 96: ......