OPERATION
Axio Examiner
Illumination and contrast procedures
Carl Zeiss
M60-2-0003 e 05/2012
65
4.1.3
Oblique illumination via condenser
(1) General operating principle
Oblique illumination allows the relief representation of your specimen. This is achieved by a radially
adjustable stop in the condenser which directs light at an angle to the specimen.
(2) Equipment
The equipment of every microscope allows the performance of the oblique illumination procedure.
−
HAL 100 halogen lamp for transmitted-light illumination
The following condensers may be used for oblique illumination:
−
Condenser 0.8 (424226-9000-000)
−
Condenser 0.9 (424226-9010-000)
−
Condenser 1.2 (424226-9020-000)
(3) Adjusting oblique illumination
•
Switch on transmitted-light illumination.
•
Set the condenser modulator disk to the
position for oblique illumination .
•
Use the adjusting screw (width A/F 1)
(Figure 4-6/
1
) at the condenser to turn the
stop's slot segment into the desired position.
The modulator disk is blocked and cannot be
turned as long as the screwdriver is inserted
for setting oblique illumination.
Figure 4-6 Adjusting oblique illumination
Содержание Axio Examiner
Страница 1: ...Operating Manual Axio Examiner Research Microscope...
Страница 13: ...INTRODUCTION Axio Examiner List of interfaces Carl Zeiss M60 2 0003 e 05 2012 13 Figure 1 2 List of interfaces...
Страница 19: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 19 2 2 System overview...
Страница 20: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 20 M60 2 0003 e 05 2012...
Страница 21: ...EQUIPMENT DESCRIPTION Axio Examiner System overview Carl Zeiss M60 2 0003 e 05 2012 21...
Страница 22: ...EQUIPMENT DESCRIPTION Carl Zeiss System overview Axio Examiner 22 M60 2 0003 e 05 2012...
Страница 96: ......