System Overview
S100 / OPMI pico
Version 7.0
Page 72
G-30-1781-en
Xenon light source
1
Ventilation grid
2
Lamp module
3
Manual switching to the backup lamp
The xenon light source has a lamp module containing two xenon lamps.
The second lamp is used as backup lamp. Manually swivel in the backup
lamp (Page 218) if it is used beyond its maximum service life of 500h (see
also chapter "Maintenance intervals" on Page 222) or if the the first lamp
fails.
If the backup lamp is operated, keep a new lamp module ready for ex-
change (Page 226).
4
Indicator: Backup lamp is in use
If the red segment (4) in the button (3) is lit, the backup lamp is operated.
5
Opening the lamp module
When you press this button, the lamp module is partially ejected.
6
Brightness control
If the light source is switched on, the brightness can be continuously set
by turning the brightness control of the fiber illumination:
CAUTION
Limited service life of the xenon lamp!
If used longer than the indicated service life of 500 hours the xenon lamp
may fail or burst.
Dropping fragments may contaminate the sterile area. Risk of infection.
• Manually swivel in the backup lamp (Page 218) or replace the lamp mod-
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...