Version 7.0
G-30-1781-en
Page 11
S100 / OPMI pico
Safety Measures
• Please take note of the planning guidelines (brochure M-30-1385) for
planning the S100 ceiling mount or the S100 wall mount.
• Note the symbols and labels attached to the device (see Page 22)!
• Use only original accessories approved by ZEISS.
Legal regulations
• Observe the legal regulations for accident prevention and occupational
health and safety applicable in the country concerned.
• This system must not be modified without the manufacturer's approval. If
the system is modified, suitable inspections and testing must be complet-
ed to ensure that it can still be used safely.
The manufacturer is not liable for damage caused by unauthorized per-
sons tampering with the system. Furthermore, this will forfeit any rights
to claim under warranty.
• Modifications and repairs on this device or any equipment operated to-
gether with this device may only be performed by ZEISS SERVICE.
• If required by the regulations and directives applicable in the country of
use, connect the system to a special emergency backup power supply.
Ambient requirements
• Do not operate the system contained in the delivery package
– in explosion-risk areas,
– if inflammable anesthetics or volatile solvents such as alcohol, benzine
or similar chemicals are present at a distance of less than 25 cm.
• Do not station or use the instrument in damp rooms. Do not expose the
instrument to water splashes, dripping water or sprayed water.
• Do not place any fluid-filled containers on top of the instrument. Make
sure that no fluids can seep into the instrument.
• Make sure that the installation conditions and the use of the device meet
surgical requirements: See Ambient conditions on Page 257
Electrical safety
• Always switch off the system before disconnecting it from / connecting it
to line power, for cleaning its surface or if it will not be used for a pro-
longed period of time.
• Switch off the unit at the power switch if you notice any smoke, sparks or
unusual noise. Do not use the unit until it has been repaired by our service
team.
• Do not use multiple sockets!
• Never open the device! The device contains freely accessible live compo-
nents. If you remove the housing, you run the risk of electric shock.
Содержание 000000-1080-349
Страница 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Страница 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Страница 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Страница 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Страница 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Страница 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Страница 111: ...Version 7 0 G 30 1781 en Page 111 S100 OPMI pico Preparations for use Fig 57 Connecting the HD video camera 4 2 1 3 5...
Страница 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Страница 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Страница 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Страница 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Страница 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Страница 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Страница 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...