3.2

 

Probe bandwidth

 

Because the 1X probe has a large input capacitance, the bandwidth of the 1X is 

usually within 5MHz; while the bandwidth of the 10X is the standard bandwidth of 
the probe.

 

3.3

 

Probe withstand voltage

 

The maximum withstand voltage of the probe is generally marked on the probe, 

which is usually 600Vpk-pk.

 

3.4

 

Use tips 

3.4.1  Large input capacitance 

Because the probe 1X has a large input capacitance (usually more than 100 pf), it 

is not suitable for measuring some capacitively sensitive circuits. For example, when 
measuring a passive crystal of MCU, the large capacitor may cause the crystal to fail 
to vibrate. As for the output of a high-speed op amp, a capacitive load may cause the 
op  amp  to  overshoot  or  oscillate.  For  circuits  with  large  capacitances  that  are  not 
suitable for measurement, please use 10X. The input capacitance of the 10X is usually 
several pf.

 

3.4.2  Measuring high voltage 

The oscilloscope's measurement voltage range is -40V ~ + 40V (80Vpp). Please 

use  the  10X  probe  to  attenuate  the  signal  by  10  times  beyond  this  range.  The 
measurable voltage range of 10X range is 220V. Please note that you must first switch 
the probe to the 10X position before connecting the high voltage signal, otherwise the 
high voltage may break down the internal circuit of the oscilloscope

3.4.3  High bandwidth 

Because the 1X probe has a large input capacitance, the bandwidth of the probe 

1X is usually within 5MHz, so please use the 10X position when the measured signal 
bandwidth is greater than 5M. 

Summary

10X is usually used in most applications. 

4

 

Vertical system 

4.1

 

Vertical sensitivity

 

  Oscilloscope vertical sensitivity refers to the voltage represented by 

a grid

 in 

the vertical direction of the screen. There are 8 divisions in the vertical direction of 

Содержание DSO1C81

Страница 1: ...DSO1C81 User s Manual DIGITAL OSCILLOSCOPE...

Страница 2: ...mal Trigger edge Rising edge trigger or falling edge trigger Battery level indication Trigger level position Timebase The time of a grid X1 X10 indicates probe attenuation which should be set to match...

Страница 3: ...ered Count turn on the pulse counter View Enter the image view page mV Save save the waveform screenshot s open the horizontal cursor ns open the vertical cursor V Zoom enter magnifying glass mode Mor...

Страница 4: ...d 2 2 Auto After the probe is connected to the signal press the Auto button and the oscilloscope will automatically detect the signal and adjust the range If there is no waveform the probe may not be...

Страница 5: ...Once the pulse crosses the trigger level the counter will be incremented by 1 Press and hold the Func key then press the up and down keys to move the trigger level 3 Probes 3 1 Probe calibration The c...

Страница 6: ...suitable for measurement please use 10X The input capacitance of the 10X is usually several pf 3 4 2 Measuring high voltage The oscilloscope s measurement voltage range is 40V 40V 80Vpp Please use the...

Страница 7: ...direction on the oscilloscope screen There are 12 grids in the horizontal direction of the oscilloscope The sampling time that can be observed on the screen time base 12 The user can estimate the sign...

Страница 8: ...is edge trigger including rising edge trigger and falling edge trigger 6 1 Trigger level In the edge trigger mode the trigger condition appears only when the waveform crosses up or down the trigger le...

Страница 9: ...rm is taken per press It is important to note that this sampling must be triggered if it is not triggered the upper left corner of the oscilloscope flashes and the sampling is not completed until ther...

Страница 10: ...the sine wave is button to turn on off the signal generator output button to reset the number to 1 00Khz 50 keys to switch the waveform type sine wave square wave can adjust the duty cycle other wavef...

Страница 11: ...he trigger condition it will automatically force the trigger to display the waveform but because there is no suitable trigger condition the waveform will not be displayed steadily For Normal mode The...

Страница 12: ...g if you need to perform a single trigger time base 200ms please turn off roll mode FFT fast Fourier transform calculation The frequency cursor is supported but the amplitude cursor is not supported 8...

Страница 13: ...480Kbit 60Kpts Frequency 0 01 Rise time 4ns Pulse counter Support Impedance 1M FFT Support Time base 5ns 10s Generator vol 3V Ver sensitivity 10mV div 10V div Generator freq 5MHz sin 1MHz other Max v...

Страница 14: ...ng to DC and adjust the probe to the 10X position the oscilloscope menu should also be set to the same 10X as the probe Connect the probe to the power output terminal and then click the Auto key Pay a...

Отзывы: