zebris Medical GmbH 

FDM Technical Data and Operating Instructions 

Page 42/54 

 

5.8 

Recommendations for recording data 

To receive significant data from FDM systems some principle guidelines should be followed. 
This chapter describes the ideal conditions for recording measurement data: The following 
points refer to the data reception of a person walking and describe the ideal measurement 
situation. 
 

5.8.1  Walking range 

The best conditions for measurements with platforms of the FDM-S type will be reached by 
integration of the system in  a walking range.  The complete walking range must  be plane 
with the surrounding floor. This 

way the test person won’t know the position of the platform 

and gives a workaround to the tendency that test persons try to walk exactly on the sensor 
area. The width of the stage should be about 1,20m. We recommend a distance of about 
4m from start to FDM-S platform and no less than 3m behind. With such a walking stage it 
is easier to measure normal walking without acceleration or deceleration. 
 
Of course the same set up can be used to measure with the method of first step. The first 
step  method  is  described  as  follow:  The  patient  stands  on  one  side  of  the  platform  in  a 
distance to reach the platform by the first step. For measurement the patient hits the sensor 
area by the first step and moves on. This kind of measurement guaranties reproducible steps 
and results. Notice: These results differ more or less from these by normal walking. 
 

5.8.2  Data recording 

Please observe the exercise of the patient strictly. Only steps where the complete ground 
contact of the foot is located on the sensor area may be used for evaluation. If not the com-
plete foot area was measured by the system (foot did partially not hit the sensor area) the 
step can not be evaluated. 
 

5.8.3  Gait velocity 

For the measurement a normal (individual) and constant walking velocity is necessary.  Ide-
ally an additionally measurement, e.g. by photo sensors, can proof the velocity for notice. 
Naturally the patients adapt to the measurement situation within a few minutes. After a few 
trials, walking seems normal. A change in velocity of about 5 % is non-effective. 
 

5.8.4  Posture 

A visual control of the behaviour pattern of normal gait is recommended. Trials with atypical 
behaviour  pattern  should  be  deleted  from  interpretation.  The  patient  has  to  look  straight 
ahead and must not be disturbed by paying attention to the platform or monitor. Marks on 
the wall in front of the patient can provide orientation to hit the platform. 
 

5.8.5  Acrosclerosis 

Different  measurements  (e.g.  P.R.  Cavanagh, 

The  Foot

  (1994) 

4

,  123-135)  show  an  in-

crease of plantar pressure peaks of about 30 % by acrosclerosis (e.g. weals). The interpre-
tation of measurement data has to include the existence of plantar acrosclerosis. 

 

 

Содержание 01243005

Страница 1: ...zebris Medical GmbH www zebris de Text Release 19 01 2016 Specifications and Operating Instructions...

Страница 2: ...TORS 18 3 5 STATUS INDICATOR LED 18 3 6 ZEBRIS SYNC 19 3 6 1 Synchronization input SYNC IN 20 3 6 2 Synchronization output SYNC OUT 21 3 6 3 Synchronizing the FDM Platform with video data Sync Audio 2...

Страница 3: ...g 46 6 4 2 Manuelle Desinfektion 46 6 5 DISPOSAL 47 6 5 1 Packaging 47 6 5 2 Disposal of electronics 47 7 SAFETY STANDARDS AND SYSTEM CLASSIFICATION 48 7 1 CLASSIFICATION ACC TO ANNEX IX OF DIRECTIVE...

Страница 4: ...me aware of any errors when using this User Manual or should you find details that do not conform with your device please kindly inform us We shall then correct any possible errors as quickly as possi...

Страница 5: ...measuring system consists of several parts 1 FDM technical specifications and user manual 2 zebris FDM user manual of the application software 3 User manual of the accessories like e g projector or PC...

Страница 6: ...irective and EMC Directive The CE mark with reference number 0535 of notified body BSI formerly EUROCAT on the type plate confirms the conformity of the system with the Directive 93 42 EEC for Medical...

Страница 7: ...ified personnel such as clinical doctors physiotherapists orthopedics which posses the ability to evaluate the output data in medical aspects as a aid for the diagnosis treatment or patient care and t...

Страница 8: ...evice can cause fire electrical shock or other severe acci dents The FDM system is NOT specified for the operation in vacuum hy perbaric or altitude chambers The measuring systems are not intended for...

Страница 9: ...ustrial safety regulations the valid accident prevention For the safety reliability and performance of the components delivered by zebris respon sibility is assumed if assembly extensions re settings...

Страница 10: ...etectable damage to the device or component parts they should be returned to the manufacturer for a safety check It is not permissible to continue using the device or its component parts as severe dam...

Страница 11: ...DM Sensor The sensors of the different FDM system only vary in size of the measuring area the number of single sensors included in the sensor module and the supported sampling frequency All other tech...

Страница 12: ...Number of sensors 40 x 48 1920 Sensor surface 400 x 330 mm L x W Resolution 1 3 resp 1 4 sensors cm Infrared interface optional Type FDM S REF No 01243010 Outer dimensions 690 x 400 x 21 mm L x W x H...

Страница 13: ...Number of sensors 60 x 64 2560 Sensor surface 810 x 510 mm L x W Resolution 1 2 resp 0 6 sensors cm Infrared interface integrated Geben Sie hier eine Formel ein Type FDM J1 8SQ REF No 01243210 Outer d...

Страница 14: ...Number of sensors 64 x 176 11264 Sensor surface 1440 x 560 mm L x W Resolution 1 3 resp 1 4 sensors cm Infrared interface integrated Type FDM 2 REF No 01243020 Outer dimensions 2122 x 605 x 21 mm L x...

Страница 15: ...pe FDM 3 REF No 01243030 Outer dimensions 3070 x 605 x 21 mm L x W x H Weight approx 35 kg Measuring frequency 100 Hz Number of sensors 64 x 352 22528 Sensor surface 2980 x 560 mm L x W Resolution 1 3...

Страница 16: ...f sensors 44 x 136 5984 Sensor surface 1730 x 560 mm L x W Resolution 1 2 resp 0 6 sensors cm Infrared interface optional Type FDM 2 4 REF No 01243035 Outer dimensions 2510 x 600 x 21 mm L x W x H Wei...

Страница 17: ...espective measuring column The analog signal coupled into the shift register over the lines is proportional to the pressure dependent capacity and is passed on for further processing to the control an...

Страница 18: ...established yet or recognized The platform is not ready for initialization or measurement green permanent The power supply unit is connected to mains and a correct sup ply voltage is provided A USB co...

Страница 19: ...1 when a third party device is synchro nized with the FDM system This allows non medical equipment to be synchronized with the FDM system as long as such devices are out of patients reach Nevertheles...

Страница 20: ...ion and pin 1 is set to 5V 1 by an internal pull up re sistor 2 7 k If this input is set to 0 V 0 i e by a switch or break contact than the SYNC IN is triggered Electrical specifications Input resista...

Страница 21: ...ill trigger a syn chronized measurement of the third party device either via start stop or frame by frame mode Electrical specifications Output resistance 100 High Level 2 0V Low Level 0 8V Respective...

Страница 22: ...as utilizing the external microphone input of the cam era The synchronization is effected by imprinting a tact signal on the soundtrack of the video recording This data is evaluated automatically by t...

Страница 23: ...in the platform housing is required FDM platform and DAB Bluetooth are synchronized automatically as soon as both devices have been switched on and a measurement is started The following schematic dia...

Страница 24: ...connected to separate USB ports of the same PC By means of the synchronization cable the Sync Out socket of the master platform has to be connected to the Sync In socket of the slave platform The Win...

Страница 25: ...e for operating system Windows 7 32 64 Bit 79010105 Hardware FDM user manual english Printing version is liable to be charged Free download of PDF Files from zebris Service Center 79010185 Software ze...

Страница 26: ...Video Sync Control Cable 2 5 Length 2 5m both sides phone jack 3 5mm without amplifier for zebris SYNCCam 21030312 Video Sync Control Extension Cable Length 5m phone jack socket 3 5mm 01540110 SYNCLig...

Страница 27: ...zebris Medical GmbH FDM Technical Data and Operating Instructions Page 27 54 01850011 SYNCLight plus Adapter Cable for Master Slave connection of up to 3 SYNCLight plus length 1m...

Страница 28: ...c LEDs are flashing when the camera is disconnected from the USB port Therefore it is strongly advised not to look directly into the camera when it is disconnected in order to avoid dazzling Technical...

Страница 29: ...en it is disconnected in order to avoid dazzling Furthermore contains the SYNCLightCam as an integral solution the LED video illumination In order to produce well lighted and tack sharp video captures...

Страница 30: ...nchronization VIDEOSYNC Ein Ausschalten mit der Messung SYNC IN Standard zebris Synchronisation kompatibel mit SYNC IN OUT Plattform Mounting Zoll Stativgewinde an der R ckseite NOTE In order to maint...

Страница 31: ...matically which results in reduced brightness in order to prevent the SYNCLight plus from being damaged by excessive heat Power Supply Unit For operation of the SYNCLight plus a power supply unit need...

Страница 32: ...tly adapted for use in combination with zebris SYNCCam as well as the force distribution measurement Their brightness can be adjusted infinitely The integrated synchronization unit automatically switc...

Страница 33: ...W x H approx 640g Power Supply 24V DC 36W Light Colour Light Current 6200K 1550 Lumen Synchronization VIDEOSYNC On Off with force measurement Mounting Inch tripod thread at back side Interpretation o...

Страница 34: ...No 01540120 Dimensions Weight 155 x 210 x 38mm L x W x H approx 640g Power Supply 24V DC 36W Light Colour Light Current 6200K 1550 Lumen Synchronization VIDEO SYNC On Off with force measurement PULSE...

Страница 35: ...VIDEO SYNC OUT and can be used for control of additional devices SYNC IN ESD protected voltage reversal proof input Input resistance 38K Pull Up VIH High Level Input Voltage 3 7V VIL Low Level Input...

Страница 36: ...SE SYNC By utilizing pulsed light optimal lighting conditions for industrial cameras can be accomplished with out being too bright or disturbing for the human eyesight In order to use PULSE SYNC the S...

Страница 37: ...an lead to inaccurate measuring results Set up the measuring system in a way that the socket for the mains connection is accessible easily at all times and that the device can be separated from the po...

Страница 38: ...cket in the connector compartment WARNING Exclusively use the power plug approved by zebris for the operation of the FDM platform which is suitable for the power supply of your platform Power Supply M...

Страница 39: ...the mains compare the nameplate indications of the power supply and the treadmill in terms of mains voltage and mains frequency with the local characteristics Only connect when they are in accordance...

Страница 40: ...nts or third parties If the zebris FDM software shall be installed in a network data network the operator is obliged to determine analyse evaluate and control the risks that are connected with doing s...

Страница 41: ...ftware Therefore please carefully read the zebris FDM user manual If the power supply unit and the USB cable are connected to the sockets of the platform please close the lid of the safety lock and fa...

Страница 42: ...nsor area by the first step and moves on This kind of measurement guaranties reproducible steps and results Notice These results differ more or less from these by normal walking 5 8 2 Data recording P...

Страница 43: ...ections and STK The zebris Medical GmbH does not stipulate any safety related control for the FDM system For maintaining the correct state of the electrical equipment checks and technical safety inspe...

Страница 44: ...uring patterns These include above all line or column shaped measuring patterns deviating from the surrounding values Whenever faults occur or in case of doubt the manufacturer or sales partner author...

Страница 45: ...se of malfunction of your FDM measuring system our service employees need the fol lowing information Device type serial number of the FDM platform Please find the serial number on the left hand side o...

Страница 46: ...tion liquid WARNING No spray disinfection Spray disinfection can destroy the highly precise measurement sensors of the platform Recommended disinfection agent Composition approx 25 ethanol 35 Propanol...

Страница 47: ...is carried out by the manufacturer For this purpose the system should be shipped to the manufacturer and will be forwarded to regular disposal by zebris The improper interaction with electronic waste...

Страница 48: ...0601 1 2006 Classification of the FDM Sensor according to DIN EN 60601 1 Type BF Safety class II Steady state conditions Unsuitable for use in an oxygen enriched atmosphere 7 2 1 Connecting the FDM Sy...

Страница 49: ...cal electrical components they may only be used outside the patient s vicinity Furthermore the user must ensure never to touch the contacts of the connectors of the interface box and the patient at th...

Страница 50: ...ly unit of the FDM sensors to the same multiple socket If multiple sockets are used jointly for components of the FDM system that are allowed to be located within the vicinity of the patient e g tread...

Страница 51: ...table and mobile RF communications equipment can affect the system If ever possible such devices should not be operated within the system environment during measurements WARNING The use of accessories...

Страница 52: ...discharge 6 kV contact discharge 8 kV atmospheric discharge Flooring should be of wood or concrete or laid with ceramic tiles If the flooring is made of synthetic material the relative humidity must b...

Страница 53: ...ansmitters If the measured field strength at the site of the FDM T force distribution measuring system exceeds the compliance levels listed above the FDM T force distribution measuring system must be...

Страница 54: ...zebris Medical GmbH FDM Technical Data and Operating Instructions Page 54 54 7 4 Declaration of conformity medical platforms...

Отзывы: