Electromagnetic interference

The EM environment for this product is the home healthcare environment and

professional healthcare facility environment.

The essential performance of this product is the accuracy of SpO

2

and pulse rate

(SpO

2

Accuracy:

±

2% in the range of 70%-100% of SpO

2

, No definition for SpO

2

under 70%;Pulse rate: 25bpm~250bpm, accuracy:

±

1% or

±

1bpm(larger)).When

used directly near strong electromagnetic interference (for example: near mobile

phones, microwave ovens, etc.), it may be temporarily inaccurate. If so, please keep

the product away from interfering devices.

The finger pulse oximeter may exhibit temporary degradation of performance

(e.g.deviation from the performance indicated in the instructions for use during

IMMUNITY testing) and the finger pulse oximeter can recover from any interference

within 30 seconds without affecting basic safety or performance without operator

intervention.The following degradations, if associated with BASIC SAFETY or

ESSENTIAL PERFORMANCE shall not be allowed:

- component failures;

- changes in programmable parameters or settings;

- reset to default settings; and

- change of operating.

During measurement, The use of this equipment adjacent to or stacked with

other equipment should be avoided because it could result in improper operation. If

such use is necessary, this equipment and the other equipment should be observed to

verify that they are operating normally.

During measurement, portable RF communications equipment (including

peripherals such as antenna cables and external antennas) should be used no closer

than 30 cm (12 inches) to any part of the finger pulse oximeter, including cables

specified by the manufacturer. Otherwise, degradation of the performance of this

equipment could result.

Test Summary

Test Reference

Standard

Test Item

Result(Pass or

Fail or N/A)

Emission Measurements

CISPR 11

Conducted Emission

N/A

CISPR 11

Radiated Emission

Pass

IEC/EN 61000-3-2

Harmonic Current Emissions

N/A

IEC/EN 61000-3-3

Voltage Fluctuations and Flicker

N/A

Содержание YX310

Страница 1: ...1 YUWELL Finger Pulse Oximeter YX310 Please read the user manual carefully before using The picture is for reference only please refer to the actual product...

Страница 2: ...ntify interference and the elimination of the following measures 1 Shut down the equipment in the vicinity and then re open in order to identify interference equipment 2 To change the direction or loc...

Страница 3: ...dex finger middle finger and ring finger instead of small finger 22 Warning Do not use the oximeter if it shows signs of damage or is suspected of being damaged Inaccurate or no readings may result fr...

Страница 4: ...wer levels 34 This product can be operated by the patient or by others to measure the patient s PR and SpO2 The maintenance operation and maintenance methods are the same This product contains batteri...

Страница 5: ...yhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology so that two beams of different wavelength of light red light and infrared light can be...

Страница 6: ...dication for Use Intended use The YUWELL Finger Pulse Oximeter is a non invasive non sterile reusable spot checking device which can measure and display SpO2 and pulse rate through finger It is intend...

Страница 7: ...approximately 2 thirds of the values were at over or below the accuracy Arms value Pulse rate 25bpm 250bpm accuracy Arms 1 or 1bpm larger 5 Anti interference ability of ambient light Deviation in blo...

Страница 8: ...P22 17 Description of oximeter application management YX310 is equipped with Bluetooth function Bluetooth communication protocol module enables the oximeter to be equipped with Bluetooth connection an...

Страница 9: ...The Skin tones included in the study were as follows 3 subjects with dark pigmentation 5 subject with very light pigmentation The remaining subjects with light medium skin tones of Asian origins 5 Gra...

Страница 10: ...all be inserted into the sensor position which is in the middle of the device The Finger Pulse Oximeter is NOT design for newborns and infants Product Operation Steps 1 Install two AAA batteries into...

Страница 11: ...the hanging hole 2 Thread thicker end of the lanyard through the threaded end before pulling it tightly 3 Install as the figures show See Figure Maintenance and Storage Under normal conditions there i...

Страница 12: ...r soft cloth moistened with water 3 After cleaning wipe off the water with a soft cloth 4 Allow the oximeter to air dry The most commonly used hospital cleaning and non corrosive liquid detergent can...

Страница 13: ...oblem exiting in the product please go to a hospital timely for exact diagnosis 3 Nail polish or paste manicure 3 Remove the nail polish or discharge manicure when measuring SpO2 or PR is shown unstab...

Страница 14: ...on The following degradations if associated with BASIC SAFETY or ESSENTIAL PERFORMANCE shall not be allowed component failures changes in programmable parameters or settings reset to default settings...

Страница 15: ...ME SYSTEMS Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING RS Guidance and manufacturer s declaration Radiated Emission Emissions test IEC60601 test level Compliance level Result...

Страница 16: ...Band 13 17 Pulse modulationb 217Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulationb 18Hz 2 0 3 28 870 930 1720 1700 19 00 GSM 1800 CDMA 1900 GSM 1900...

Страница 17: ...ent The device has been evaluated to meet general RF exposure requirement The device can be used in portable exposure condition without restriction Note This equipment has been tested and found to com...

Страница 18: ...Card Thank you very much for using our products Product name Finger pulse oximeter Model YX310 S N MFG DATE JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Manufacturer Address No 1 Baisheng Road Develo...