Physical properties
Dimensions
(length*width*height)
270mm * 138mm * 100mm or
10.63'' * 5.43'' * 3.94''
Weight
Air tube
Plastic hose,about 1.8 m
Maximum volume of
water tank
260±10 mL
Air outlet
22 mm (complies with ISO 5356-1:2015)
Temperature
Maximum temperature
of heater plate
55
℃ ±4℃ ( 131℉ ±14.
4
℉
)
Cut-out
110℃(if it is damaged, return to manufacturer)
Maximum gas
temperature
43℃
Material: Polyurethane
Average arrestance: 80% for~10 micron dust
Therapy pressure
4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH
2
O
Initial pressure
4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH
2
O
Maximum pressure
4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH
2
O
Minimum pressure
4~20 hPa ( ≤ maximum pressure, adjustable, step is 0.5 hPa) or 4~20 cmH
2
O
Ramp
0~45 minutes (adjustable, step is 5 minutes)
Gas leakage
2 L/min at the maximum rated pressure
The performance of the device at set pressure according to ISO 80601-2-70:2015
is shown below:
Test pressures(hPa)
4
8
12
16
20
CONNECTION PORT (L/min)
≥ 100
≥ 100
≥ 100
≥ 100
≥ 100
performance
74:2017)
Relative humidity: ≥ 50%
Pressure accuracy
Maximum static pressure variation according to ISO 80601-2- 70:2015
±[2% of the full scale +4% of the set value]
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Pressure(cmH
2
O)
10 bpm
15 bpm
20 bpm
4
± [2% of the full scale +4% of the set value]
8
± [2% of the full scale +4% of the set value]
12
± [2% of the full scale +4% of the set value]
16
± [2% of the full scale +4% of the set value]
20
± [2% of the full scale +4% of the set value]
Maximum limited
pressure
The maximum limited pressure
20 cmH
2
O in normal condition
40 cmH
2
O under single fault Condition
The Uncertainty of
measurement:
For measures of pressure: ±0.2 hPa
Atmosphere
Filter
Blower
Air tube
Leak
Mask
Flow
sensor
Pressure
sensor
Water
tank