Physical properties

Dimensions 
(length*width*height) 

270mm  *  138mm  *  100mm or 
10.63''  *  5.43''  *  3.94'' 

Weight 
Air tube 

Plastic hose,about 1.8 m

Maximum volume of 
water tank

260±10 mL 

Air outlet

22 mm (complies with ISO 5356-1:2015)

Temperature

Maximum temperature 
of heater plate 

55

℃ ±4℃ ( 131℉ ±14.

4

)

Cut-out

110℃(if it is damaged, return to manufacturer)

Maximum gas 
temperature

43℃

Material: Polyurethane 
Average arrestance: 80% for~10 micron dust

Therapy pressure 

4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Initial pressure

4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Maximum pressure 

4~20 hPa (adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Minimum pressure 

4~20 hPa ( ≤ maximum pressure, adjustable, step is 0.5 hPa) or 4~20 cmH

2

O

Ramp

0~45 minutes (adjustable, step is 5 minutes) 

Gas leakage 

2 L/min at the maximum rated pressure

The performance of the device at set pressure according to ISO 80601-2-70:2015 
is shown below:

Test pressures(hPa)

4

12

16

20

CONNECTION PORT (L/min)

≥ 100

≥ 100

≥ 100

≥ 100

≥ 100

performance 

74:2017)
Relative humidity: ≥ 50%

Pressure accuracy

Maximum static pressure variation according to ISO 80601-2- 70:2015 
±[2% of the full scale +4% of the set value]

Maximum dynamic pressure variation according to ISO 80601-2-70:2015 

Pressure(cmH

2

O)

10 bpm

15 bpm

20 bpm 

4

± [2% of the full scale +4% of the set value]

8

± [2% of the full scale +4% of the set value]

12

± [2% of the full scale +4% of the set value]

16

± [2% of the full scale +4% of the set value]

20

± [2% of the full scale +4% of the set value]

Maximum limited 
pressure 

The maximum limited pressure 
20 cmH

2

O in normal condition 

40 cmH

2

O under single fault Condition

The Uncertainty of 
measurement:

For measures of pressure: ±0.2 hPa

Atmosphere

Filter

Blower

Air tube

Leak

Mask

Flow

sensor

Pressure

sensor

Water

tank

Содержание BreathCare PAP

Страница 1: ...zhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 Web address www yuyue com cn Shanghai International Holding Corp GmbH Europe IFU V 02 version 04 Prepared date 2022 06 Breat...

Страница 2: ...f a physician 1 Indications for use 01 2 Contraindications 01 3 Warnings 01 4 Precautions 03 5 Adverse reactions 04 6 Package table 05 7 Picture and explanation for product 05 8 Explanation of button...

Страница 3: ...the device where it can be bumped or where someone is likely to trip over the power cord Do not block the air tube and or air inlet of the device while in operation otherwise it could cause overheatin...

Страница 4: ...age the device Use only vented masks recommended by Yuwell or by your doctor with this device Fitting the mask without the device blowing air can result in rebreathing of exhaled air Make sure that th...

Страница 5: ...tion of button START STOP button Press to start stop therapy Dial Turn to navigate the menu and press to select an option Turn to adjust the options and press to save your choice 9 Installation The fo...

Страница 6: ...utomatically after two minutes of no action You can press any button to turn it back on Caution When the level of the humidity is 1 6 and the device is delivering gas pressure the heating plate will h...

Страница 7: ...6 55 This temperature has been tested with empty water tank with 10 minutes preheating time To adjust the humidity level Highlight the Humidity press dial to enter the humidity page Press the dial and...

Страница 8: ...se or decrease therapy pressure each step is 0 5 cmH2O hPa After setting therapy pressure press dial to save your change Highlight the Back press dial to back to home page You can set Max Pressure Min...

Страница 9: ...t Notice Don t pull the threaded portion of the air tubing 12 2 Cleaning WARNINGS Regular cleaning of the device and its accessories is very important for the prevention of respiratory infections To a...

Страница 10: ...daily cleaning of the water chamber 12 2 2 Cleaning the Enclosure Wipe the surface of the device with a soft slightly damp cloth CAUTIONS The device can only be used after the enclosure is dry so tha...

Страница 11: ...f the mask is not suitable See your mask user guide to check Please refer to the nose measurement card to choose the appropriate mask size I am getting a dry or blocked nose Humidity level may set too...

Страница 12: ...contact your provider or Yuwell ERROR 8 There is a mistake of blower Please contact your provider or Yuwell ERROR 9 There is a mistake of RTC loss power Please adjust the time of the device or contac...

Страница 13: ...leakage 2 L min at the maximum rated pressure The performance of the device at set pressure according to ISO 80601 2 70 2015 is shown below Test pressures hPa 4 8 12 16 20 CONNECTION PORT L min 100 1...

Страница 14: ...d up to 15 degrees 17 2 Disposal statement Warning Please contact local authorities your provider or Yuwell to determine the proper method of disposal of this device 18 Limited warranty Yuwell warrant...

Страница 15: ...s Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle 70 UT 25 30 cycle At 0 0 UT 250 300...

Страница 16: ...PMENT or ME SYSTEM may be reduced to 1 m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 du...

Страница 17: ...gn Date Thislimitedwarrantydoesnotcover Any damage caused as a result of improper use abuse modi cation or alteration of the product Repairs carried out by any service organizations that have not been...

Страница 18: ......

Отзывы: