Table 4 

Recommended separation distances

Device

 is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The customers or the users of these YUWELL Nebulizers 
can  help  prevent  electromagnetic  interference  by  maintaining  a  minimum  distance 
between  portable  and  mobile  RF  communications  equipment  (transmitters)  and  The 
involved model

 

as  recommended  below,  according  to  the  maximum  output 

power  of  the  communications equipment. 

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in meters (m) can be estimated using the equation applicable to 
the  frequency  of  the  transmitter,  where  P  is  the  maximum  output  power  rating  of  the 
transmitter in watts (W) according to the transmitter manufacturer. 

Note: At 80MHz and 800MHz, the separation distance for the higher frequency range 
applies. 

Note: These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects, and people.

Output Power of

Transmitter in

Watt

Separation distance according to frequency 

of transmitter in meter

150kHz

80MHz

d=1.2

80MHz

800MHz

d=1.2

800MHz ~ 2.5GHz

d=2.3

P

P

P

0.01                             0.12                             0.12                              0.23

0.1                               0.38                             0.38                              0.73

 1                                  1.2                              1.2                                2.3

10                                 3.8                              3.8                                7.3

100                               12                               12                                 23

Recommended separation distance between portable and mobile RF communications 
equipment and the involved model

Immunity test

IEC 60601 

Test level

Compliance 

level

Electromagnetic environment 

– guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 V rms
150 kHz ~ 
80 MHz

3 V/m
80 MHZ to
2.5 GHz

3 V rms

3 V/m

Portable and mobile RF communications 
equipment  should  be  used  no  closer  to 
any  part  of  the  YUWEII  nebulizer 

 

including cables, than the 

recommended separation  distance 

calculated  from  the  equation appropriate 

to the frequency of  the transmitter. 

Recommend separation distance 

d=1.2
d=1.2      80MHz to 800 MHz
d=2.3      800 MHz to 2.5 GHz
where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the recommended 
separation distance in meters (m). 
Field strengths from fixed RF 
transmitters as determined by an 
electromagnetic site survey, a should be 
less than the compliance level in each 
frequency range. b Interference may 
occur in the vicinity of equipment marked 
with he following symbol:

Note1: At 80 MHz and 800 MHz, the higher frequency range applies. 
Note2: These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is 
affected by absorption and reflection from structures, objects, and people. 

a)Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/ 
cordless)  telephones  and  land  mobile  radio,  AM  and  FM  radio  broadcast,  and  TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment  due  to  fixed  RF  transmitters,  an  electromagnetic  site  survey  should  be 
considered.  If  the  measured  field  strength  in the location in which involvedmodel

  is used 

exceeds  the  applicable  RF  compliance  level  above,  the  model should  be  observed  to 
verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures  may 
be necessary, such as reorienting or relocating the Nebulizer. 
b)Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m.

P

P

P

Guidance and manufacturer's declaration – electromagnetic immunity 

Device

 is intended for use in the electromagnetic environment specified below. The customer 

or the user of these YUWEII Nebulizers should assure that it is used in such environment.

Table 3 

Guidance and manufacturer's declaration – electromagnetic immunity