average replacement interval is one to two weeks. Probes in
constant or heavy use may require more frequent membrane
changes.

Interferences

Hydrogen sulfide, sulfur dioxide, halogens, carbon monoxide,
chlorine, nitric oxide, and nitrous oxide can cause the probe to
give erroneous readings. If you suspect erroneous readings, it
may be necessary to determine if these are the cause.

Acids

 Avoid any environment that contains substances such as
concentrated acids, caustics, and strong solvents that may
attack the probe. Probe materials that may be damaged by
these substances include FEP Teflon, EPR rubber, ABS
plastic, and stainless steel.

Erroneous Readings

Erroneous readings may occur if the membrane is coated with
oxygen consuming bacteria or oxygen evolving algae. Heavy
residue may coat the membrane causing incorrect readings.
Frequent membrane changes will eliminate this problem.

Storage

When the probe is not in use, store the probe in a BOD bottle
containing at least 1 inch of water. For long term storage,
remove the membrane cap, rinse the probe tip with deionized
water, and install a dry membrane cap (without electrolyte
solution).

Содержание YSI 5905

Страница 1: ...YSI 5905 5010 BOD Probe Instruction Manual...

Страница 2: ......

Страница 3: ...HANGING MEMBRANE CAPS FIGURE 2 3 WARM UP 5 ELECTRODE CLEANING FIGURE 3 5 OPERATING PROCEDURES AND PRECAUTIONS 7 PROBLEM SOLVING 9 TROUBLE SHOOTING GUIDE 9 SERVICING 11 WARRANTY AND REPAIR 13 CLEANING...

Страница 4: ...ii...

Страница 5: ...meters except for the Model 55 or Model 95 It must be used with an adapter 5011 in order to be used on the Model 5000 and 5100 The 5010 probe is specifically designed for use with the Model 5000 and 5...

Страница 6: ...ectrolytic cell and acts as a reference electrode Upon entering the cell through the membrane oxygen is reduced at an applied potential of 0 8 V referenced to the silver electrode The reduction curren...

Страница 7: ...aight out 2 Unscrew the old membrane cap from the probe Before installing a new membrane clean the probe tip with deionized water in order to remove any contaminants 3 Hold the membrane cap and fill i...

Страница 8: ...4 Fill Membrane Cap with Solution Unscrew Cap Remove Stir Paddle Screw Cap on moderately tight Replace Stir Paddle Figure 2 Membrane Installation on the 5905 5010 Probe...

Страница 9: ...gold cathode must be textured properly It can become tarnished or plated with silver after extended use The gold cathode can be cleaned by using the adhesive backed sanding disc provided in the 5906 M...

Страница 10: ...Silver Anode It is normal for a dark layer of silver chloride to cover the silver anode After prolonged use it may become necessary to clean the anode Soak the probe in a 14 ammonium hydroxide solutio...

Страница 11: ...ling through it e g nitrogen argon The meter should read less than 1 dissolved oxygen in either of these environments If it does not change the membrane or clean the probe Membrane Life Membrane life...

Страница 12: ...and strong solvents that may attack the probe Probe materials that may be damaged by these substances include FEP Teflon EPR rubber ABS plastic and stainless steel Erroneous Readings Erroneous reading...

Страница 13: ...ibrate Allow approximately 30 min for probe to polarize when first turned on Change membrane Clean probe see instructions under Electrode Cleaning Does not return to calibration value Allow sufficient...

Страница 14: ...ange membrane Clean probe see instructions under Electrode Cleaning Motor not working Check power supply Manually turn the stir paddle to help start the motor Replace motor Loud motor Check if stir pa...

Страница 15: ...he following steps 1 Remove stir paddle by pulling it straight out 2 Remove the two screws A on the outside of the housing using a 3 32 Allen wrench 3 Loosen the top nut of the cable strain relief B f...

Страница 16: ...12 Housing Assembly Motor Assembly Probe Assembly Stir Paddle A B A D D C C Figure 4 Exploded View of 5905 5010 Probe...

Страница 17: ...placement Limitation of Warranty This Warranty does not apply to any YSI product damage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructio...

Страница 18: ...CE CENTERS For the nearest authorized service center contact YSI Technical Support 1725 Brannum Lane Yellow Springs Ohio 45387 Phone 1 937 767 7241 800 897 4151 US Fax 1 937 767 9353 E Mail environmen...

Страница 19: ...withhold service until appropriate cleaning decontamination and certification has been completed We will contact the sender for instructions as to the disposition of the equipment Disposition costs wi...

Страница 20: ...age 5 Cleaning must be completed and certified on any product before returning it to YSI PACKING INSTRUCTIONS 1 Clean and decontaminate items to insure the safety of the handler 2 Complete and include...

Страница 21: ..._______________ State ______ Zip ________ Country __________________ Model No of Device ______ Lot Number _________ Contaminant if known _____________________ Cleaning Agent s used ___________________...

Страница 22: ...solved Oxygen Accuracy 0 1 mg L or 2 of reading whichever is greater Operating Temperature Range 15 to 35 C Temperature Accuracy 0 2 C Polarizing Voltage 0 8 volts Response Time Typical response to a...

Страница 23: ...cludes an adapter to fit 5 taper per side smaller bottles with built in overflow funnels 5905 K 5010 K Includes a s 19 funnel 5905 L 5010 L Includes a s 24 funnel 5905 W 5010 W 6 long probe for use wi...

Страница 24: ...10593 110592 110591 N A 073035 073047 005011 5905 L 059880 059902 115904 110594 110591 059911 073035 073047 005011 5905 K 059880 059902 115904 110594 110591 059933 See dealer See dealer 005011 5905 J...

Страница 25: ...1725 Brannum Lane Yellow Springs Ohio 45387 USA 059882 Web 937 767 7241 800 765 4974 A59882H Fax 937 767 9353 September 99 Info YSI com www YSI com 1997 YSI Incorporated...

Отзывы: