11
S
ERVICING
The 5905/5010 has four user replaceable parts. The motor
assembly can be replaced if it becomes noisy, unbalanced, or
malfunctions. The probe assembly, housing assembly, and stir
paddle can also be replaced if they become damaged.
To disassemble the probe perform the following steps:
1.
Remove stir paddle by pulling it straight out.
2.
Remove the two screws (A) on the outside of the housing
using a 3/32 Allen wrench.
3.
Loosen the top nut of the cable strain relief (B) from the
housing. This will allow the cable to feed through the
housing assembly.
4.
Pull the probe and motor assembly out of the housing and
disconnect the two plugs (C).
5.
To remove the motor assembly from the probe assembly,
remove the two screws (D) using the 3/32 Allen wrench.