conductIvIty unIts (cond. unIts)

Highlight Cond. Units (Conductivity Units) and press enter to open a submenu 

that allows you to select the conductivity units to be displayed on the Run screen. 

Highlight a unit and press enter to enable or disable it. An enabled conductivity 

unit  will  have  an  ‘X’  in  the  box  next  to  it.  Highlight  the  ESC-Exit  box  along 

the bottom of the display and press enter to save any changes and to close the 

conductivity units submenu. 

There  are  seven  options  for  displaying  conductivity.  only  four  units  can  be 

enabled at the same time:

CoND-mS/cm displays conductivity in milliSiemens per centimeter.

• 

CoND-uS/cm displays conductivity in microSiemens per centimeter.

SPC-mS/cm  displays  Specific  Conductance  in  milliSiemens  per

centimeter.  Specific  Conductance  is  temperature  compensated

conductivity.

SPC-uS/cm  displays  Specific  Conductance  in  microSiemens  per

centimeter.  Specific  Conductance  is  temperature  compensated

conductivity.

Sal ppt displays salinity in parts per thousand. The salinity reading is

calculated from the instrument’s conductivity and temperature values

using  algorithms  found  in 

Standard  Methods  for  the  Examination  of

Water and Wastewater

.

TDS  g/L  displays 

Total  Dissolved  Solids  in  grams  per  liter.  TDS  is

calculated from conductivity and temperature using a user-selectable

TDS constant.

TDS mg/L displays 

Total Dissolved Solids in milligrams per liter. TDS

is calculated from conductivity and temperature using a user-selectable

TDS constant.

Note:  1 milliSiemen = 1,000 microSiemens. 

SPECIFIC CoNDUCTANCE

The conductivity of a sample is highly dependent on temperature, varying as 

much as 3% for each change of one degree Celsius (temperature coefficient = 

3%/°C). In addition, the temperature coefficient itself varies with the nature of 

the ionic species present in the sample. Therefore, it is useful to compensate for 

this temperature dependence in order to quickly compare conductivity readings 

taken at different temperatures. 

The  Pro30  can  display  non-temperature  compensated  conductivity  as  well  as 

temperature  compensated  Specific  Conductance.  If  Specific  Conductance  is 

selected, the Pro30 uses the temperature and conductivity values associated with 

each measurement to calculate a specific conductance value compensated to a 

user selected reference temperature, see below. Additionally, the user can select 

the temperature coefficient from 0% to 4%.  

Using the Pro30’s default reference temperature and temperature coefficient (25 

°C and 1.91%), the calculation is carried out as follows:

Specific Conductance (25°C) = Conductivity of sample

           1 + 0.0191 * (T - 25)

T = Temperature of the sample in °C

conductIvIty auto staBle (cond. auto staBle)

Auto Stable utilizes preset values to indicate when a reading is stable. The preset 

values are adjustable in the System Setup menu. The user can input a % change 

in readings (0.0 to 1.9) over ‘x’ amount of time in seconds (3-19). 

Highlight Cond. Auto Stable, then press enter to open the submenu. 

Use the up or down arrow keys to highlight the % change or seconds (secs) input 

field, then press enter to make the highlighted field adjustable. Use the up or 

down arrow keys to adjust the selected value, then press enter to confirm changes.  

once you have confirmed any changes, highlight the ESC-Exit box along the 

bottom of the display and press enter to close the Auto Stable submenu.  

To disable Auto Stable, set the % Change input to 0.0. 

When Auto Stable is enabled, an

AS

 symbol will display next to the reading 

on the Run screen and blink during stabilization. When the dissolved oxygen 

and/or  conductivity  reading  stabilizes  based  on  the  Auto  Stable  settings,  the

AS

 

symbol will display steadily and the instrument will beep twice if Audio 

is turned on. 

temPerature unIts

Highlight  Temperature  Units  and  press  enter  to  open  a  submenu  that  allows 

you  to  change  the  temperature  units  displayed  on  the  Run  screen.  Highlight 

the desired unit (Celsius or Fahrenheit) and press enter to enable. The enabled 

temperature unit will have an ‘X’ in the box next to it. only one unit may be 

enabled at a time. Highlight the ESC-Exit box and press enter to save any changes 

and to close the Temperature Units submenu.

Содержание Pro30

Страница 1: ...USER MANUAL English Pro30 ...

Страница 2: ...er On 5 System Setup Menu 6 Audio 7 Contrast 7 Conductivity Units Cond Units 8 Conductivity Auto Stable Cond Auto Stable 9 Temperature Units 9 Specific Conductance Reference Temperature SPC Ref Temp 10 Specific Conductance Temperature Coefficient SPC C 10 TDS Constant 10 Language 11 Auto Shutoff 11 Cell Constant 11 Resetting the System Setup Menu to Factory Default 12 Calibration 12 Temperature 12...

Страница 3: ...1 937 767 7241 800 897 4151 or visit www YSI com for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the original warranty period or at least 90 days from date of repair...

Страница 4: ...parts with items on the packing list If any parts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased Battery Installation The instrument requires 2 alkaline C cell batteries Under normal conditions the average battery life is 425 hours at room temperature without using the back lig...

Страница 5: ...long the bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Press and hold for 3 seconds to turn instrument off 4 Menu Use to enter the System Setup menu from the Run screen 5 Enter Press to confirm entries and selections 6 Down Arrow Use to navigate throu...

Страница 6: ... for three seconds Navigation The up and down arrow keys allow you to navigate through the functions of the Pro30 Navigating the Run Screen When in the Run screen the up and down arrow keys will move the highlighted box along the bottom options Once a box is highlighted press enter to access the highlighted option Description of Run screen box functions from left to right Option Description SAVE H...

Страница 7: ... select Language Press to confirm Figure 6 page 1 of System Setup menu Audio is enabled Audio Audio can be enabled or disabled by using the up or down arrow keys to highlight Audio and pressing enter When enabled there will be an X in the box next to Audio When Audio is enabled the Pro30 will beep twice to indicate stability when Auto Stable is enabled The instrument will also beep when a key is p...

Страница 8: ...ro30 can display non temperature compensated conductivity as well as temperature compensated Specific Conductance If Specific Conductance is selected the Pro30 uses the temperature and conductivity values associated with each measurement to calculate a specific conductance value compensated to a user selected reference temperature see below Additionally the user can select the temperature coeffici...

Страница 9: ...he remaining solid 4 Divide the weight of the solid in grams by the volume of water used 5 in liters to yield the TDS value in g L for this site Divide the TDS value in g L by the specific conductance of the water in 6 mS cm to yield the conversion multiplier Be certain to use the correct units If the nature of the ionic species at the site changes between sampling studies the TDS values will be i...

Страница 10: ...e When calibrating Specific Conductance the Pro30 uses the factory default values for the Specific Conductance Reference Temperature and the Specific Conductance Temperature Coefficient regardless of what is configured in the System Setup Menu The default value for the Reference Temperature is 25 C and the default value for the Temperature Coefficient is 1 91 C It is important to note that the Tem...

Страница 11: ...nsure the entire conductivity sensor is submerged in the solution or the instrument will read approximately half the expected value Gently move the probe up and down to remove any air bubbles from the conductivity sensor Turn the instrument on and allow the conductivity and temperature 2 readings to stabilize Press and hold the Cal key for 3 seconds Highlight Conductivity and press enter Next high...

Страница 12: ... display are different in Data mode figure 9 ThePro30cannotcommunicatetoaPCviaaProSeriescommunications saddle Connecting the Pro30 to a communication saddle may cause erratic instrument behavior i Figure 9 Data mode Viewing Data Once in Data mode use the up and down arrow keys to view saved data sets in sequential order or press enter to access the bottom functions After accessing the bottom funct...

Страница 13: ...instrument utilizes a gasket as a seal to prevent water from entering the battery compartment Following the recommended procedures will help keep the instrument functioning properly If the gasket and sealing surfaces are not maintained properly it is possible that water can enter the battery compartment If water enters this area it can severely damage the battery terminals causing loss of battery ...

Страница 14: ...les near the cable see figure 8 Verify the calibration solution 3 is not expired or contaminated Try a new bottle of solution Ensure you are entering in the correct 4 value for the solution according to the measurement units 1 mS 1 000 uS Allow sufficient stabilization time 5 for conductivity and temperature AND wait at least 3 seconds before confirming a calibration Contact YSI Tech Support 6 Sym...

Страница 15: ...ranging 0 0001 to 0 1 mS cm 0 1 to 0 uS cm range dependent Instrument only 0 5 of the reading or 1 uS cm whichever is greater Instrument with 1 or 4 meter cables 1 0 of the reading or 1 uS cm whichever is greater Instrument with 10 20 or 30 meter cables 2 0 of the reading or 1 uS cm whichever is greater Salinity 0 to 70 ppt 0 1 ppt 1 0 of the reading or 0 1 ppt whichever is greater Total Dissolved...

Отзывы: