CALIBRATING IN MG/L oR PPM    

Power the instrument on and place the sensor into a sample that has 

1. 

been  titrated  to  determine  the  dissolved  oxygen  concentration.  The 

sample should cover the two holes of the conductivity sensor that are 

located near the cable, see figure 8. During the calibration, continuously 

stir or move the probe through the sample at a rate of at least 6 inches 

(15.5 cm) per second if using a yellow membrane or at least 3 inches 

(7.7  cm)  per  second  if  using  a  blue  membrane.    A  stir  plate  may  be 

helpful for this calibration procedure.  

Allow the dissolved oxygen and temperature readings to stabilize. This 

2. 

may take 5 to 15 minutes, depending on the type and condition of the 

sensor.    

Press the Cal key.  Highlight Dissolved oxygen and press enter. 

3. 

Highlight mg/L or ppm depending on what is enabled in the System 

4. 

Setup menu and press enter.  

Use the up and down arrow keys to adjust the mg/L (ppm) reading to 

5. 

the value of the titrated sample. Press enter to confirm the value and 

complete the calibration or press the Cal key to cancel the calibration.

‘Calibration  Successful’  will  display  for  a  few  seconds  to  indicate  a 

6. 

successful calibration and then the instrument will return to the Run 

screen.

If the calibration is unsuccessful, an error message will display on the 

7. 

screen.  Press  the  Cal  key  to  exit  the  calibration  error  message  and 

return to the Run screen. See the Troubleshooting guide for possible 

solutions.  

conductIvIty calIBratIon

Ensure the conductivity sensor is clean and dry before performing a conductivity, 

specific conductance or salinity calibration. 

CALIBRATING 

SPECIFIC 

(SP.) 

CoNDUCTANCE 

oR 

CoNDUCTIVITY

Note:    When  calibrating  Specific  Conductance,  the  Pro2030  uses  the  factory 

default  values  for  the  Specific  Conductance  Reference  Temperature  and  the 

Specific Conductance Temperature Coefficient regardless of what is configured in 

the System Setup Menu. The default value for the Reference Temperature is 25°C 

It  is  not  necessary  to  calibrate  conductivity,  specific  conductance 

and salinity. Calibrating one of these parameters will simultaneously 

calibrate the others. YSI recommends calibrating specific conductance 

for greatest ease.

i

and the default value for the Temperature Coefficient is 1.91%/°C. It is important 

to note that the Temperature Coefficient of a calibration solution is dependent 

on the contents of the solution. Therefore, YSI recommends using a traceable 

calibration solution made of KCl (potassium chloride) when calibrating Specific 

Conductance since these solutions typically have a Temperature Coefficient of 

1.91%/°C. Additionally, be sure to enter the value of the solution as it is listed for 

25°C when calibrating Specific Conductance.  

Fill a clean container (i.e. plastic cup or glass beaker) with fresh, traceable 

1. 

conductivity calibration solution and place the sensor into the solution. 

The solution must cover the holes of the conductivity sensor that are 

closest to the cable (figure 8). Ensure the entire conductivity sensor is 

submerged in the solution or the instrument will read approximately  

half the expected value. Gently move the probe up and down to remove 

any air bubbles from the conductivity sensor.

Figure 8, solution above two holes near cable.  

Turn the instrument on and allow the conductivity and temperature 

2. 

readings to stabilize. Press the Cal key. Highlight Conductivity and press 

enter. Next, highlight the desired calibration method, Sp. Conductance 

or Conductivity, and press enter. 

Highlight the units you wish to calibrate, either uS/cm or mS/cm, and 

3. 

press enter. 1 mS = 1,000 uS. Next, use the up or down arrow key to 

adjust the value on the display to match the value of the conductivity 

calibration solution. If calibrating conductivity, it is necessary to look 

Содержание Pro2030

Страница 1: ...USER MANUAL English Pro2030 ...

Страница 2: ...Navigation 7 First Power On 8 System Setup Menu 8 DO Local 9 Last Digit Suppression LDS 10 Quick DO Calibration Quick DO Cal 10 Audio 11 Contrast 11 Sensor Type 11 Membrane Type 12 Auto Stable 13 DO Units 14 Conductivity Units Cond Units 14 Specific Conductance Reference Temperature SPC Ref Temp 15 Specific Conductance Temperature Coefficient SPC C 16 TDS Constant 16 Temperature Units 17 Pressure ...

Страница 3: ...lace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit www YSI com for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selec...

Страница 4: ...parts with items on the packing list If any parts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased Battery Installation The instrument requires 2 alkaline C cell batteries Under normal conditions the average battery life is 425 hours at room temperature without using the back lig...

Страница 5: ...turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Press and hold for 3 seconds to turn instrument off 4 Menu Use to enter the System Setup menu from the Run screen 5 Enter Press to confirm entries and selections 6 Down Arrow Use to navigate through menus to navigate through box options at the bottom of the Run screen and to decrease numerical inpu...

Страница 6: ...head connector on the cable are 3 clean and dry Grasp the sensor with one hand and the cable bulkhead in the other 4 Push the sensor into the connector on the cable until it is firmly seated and 5 only 1 o ring is visible Failure to properly seat the sensor may result in damage Twist the sensor clockwise to engage the threads and finger tighten Do 6 NOT use a tool This connection is water tight Fo...

Страница 7: ...information before displaying the main Run screen figure 5 The first time the Pro2030 is turned on it will step through language sensor and membrane selections see the First Power On section of this manual for more information Figure 5 Run screen Backlight Once the instrument is powered on pressing the power backlight key will turn on the display backlight The backlight will remain on until the ke...

Страница 8: ...tup menu contains multiple screens that are notated as pages The current page is indicated near the bottom of the display figure 7 Use the up and down arrow keys to scroll through menu options and menu pages Exiting the System Setup Menu To exit the System Setup menu press the down arrow key until the ESC Exit box is highlighted then press enter to return to the Run screen DO Local DO Local can be...

Страница 9: ... For more information on Quick DO Cal see the Calibration section of this manual Audio Audio can be enabled or disabled by using the up or down arrow keys to highlight Audio and pressing enter When enabled there will be an X in the box next to Audio When Audio is enabled the Pro2030 will beep twice to indicate stability when Auto Stable is enabled The instrument will also beep when a key is presse...

Страница 10: ...itional Teflon membranes 5914 Blue 2 0 mil polyethylene PE Less flow dependence than 1 25 mil membrane but somewhat slower response Polarographic Membrane Kits Item Color Material Description 5908 Yellow 1 25 mil polyethylene PE Faster response time and less flow dependence than traditional Teflon membranes 5909 Blue 2 0 mil polyethylene PE Less flow dependence than 1 25 mil membrane but somewhat ...

Страница 11: ... temperature values using algorithms found in Standard Methods for the Examination of Water and Wastewater TDS g L displays Total Dissolved Solids in grams per liter TDS is calculated from conductivity and temperature using a user selectable TDS constant TDS mg L displays Total Dissolved Solids in milligrams per liter TDS is calculated from conductivity and temperature using a user selectable TDS ...

Страница 12: ...ess enter to open the submenu With the TDS Constant highlighted press enter to make the field adjustable Next use the up or down arrow key to increase or decrease the value The input range is 0 30 to 1 00 Press enter to save the new TDS Constant Next highlight the ESC Exit box and press enter to close the submenu Temperature Units Highlight Temperature Units and press enter to open a submenu that ...

Страница 13: ... the Pro2030 is calibrated at the factory The barometer reading must be accurate to ensure accurate DO calibrations and readings If the barometer requires an adjustment use the up or down arrow keys to highlight the barometer box along the bottom of the Run screen then press enter Next use the up or down arrow keys to adjust the barometer reading to the local true barometric pressure Continually d...

Страница 14: ...necessary to calibrate in both and mg L or ppm Calibrating in will simultaneously calibrate mg L and ppm and vice versa YSI recommends calibrating the dissolved oxygen sensor in for both ease and accuracy i Calibrating in Percent DO Perform this calibration procedure when Quick DO Cal is disabled in the System Setup menu Perform steps 1 4 of the Quick DO Calibration procedure 1 Press and hold the ...

Страница 15: ...e Reference Temperature and the SpecificConductanceTemperatureCoefficientregardlessofwhatisconfiguredin the System Setup Menu The default value for the Reference Temperature is 25 C It is not necessary to calibrate conductivity specific conductance and salinity Calibrating one of these parameters will simultaneously calibrate the others YSI recommends calibrating specific conductance for greatest ...

Страница 16: ...ey to exit the calibration error message and return to the Run screen See the Troubleshooting guide for possible solutions T aking M easurements Before taking measurements be sure the instrument has been calibrated to ensure the most accurate readings Turn the instrument on and wait 5 15 minutes if using a polarographic sensor Install the sensor guard to protect the sensor and membrane Place the p...

Страница 17: ...to access Data mode Note that the function boxes at the bottom of the display are different in Data mode figure 10 Figure 10 Data mode Viewing Data Once in Data mode use the up and down arrow keys to view saved data sets in sequential order or press enter to access the bottom functions After accessing the bottom functions highlight the Data box and press enter to regain access to viewing data The ...

Страница 18: ...ng properly If the gasket o rings and sealing surfaces are not maintained properly it is possible that water can enter the battery compartment and or sensor port of the instrument If water enters these areas it can severely damage the battery terminalsorsensorportcausinglossofbatterypower falsereadingsandcorrosion to the sensors or battery terminals Therefore when the battery compartment lid is re...

Страница 19: ...over grease the o rings The excess grease may collect grit particles that can compromise the seal Excess grease can also cause the waterproofing capabilities of the o ring to diminish potentially causing leaks If excess grease is present remove it using a lens cloth or lint free cloth i i Sensor Maintenance Conductivity The openings that allow sample access to the conductivity electrodes should be...

Страница 20: ...sensor tip directly down on the sanding disc and twist it in a circular motion to sand the gold cathode The goal is to sand off any build up and to lightly scratch the cathode to provide a larger surface area for the electrolyte solution under the membrane Usually 3 to 4 twists of the sensor are sufficient to remove deposits and for the gold to appear to have a matte finish Rinse the sensor thorou...

Страница 21: ...ranes Chemical cleaning should be performed as infrequently as possible First attempt a membrane change and recalibrate If a new membrane does not resolve the problem then proceed with cleaning i At the time the cap membrane is changed YSI recommends that you rinse the anode and cathode figure 12 with distilled or deionized water and wipe with a clean paper towel If white deposits are evident on t...

Страница 22: ... membrane 2 type selection in the System Setup menu Calibration sleeve may not be 3 100 water saturated ensure sponge is moistened Ensure adequate sample movement if 4 performing mg L or ppm calibration Allow sufficient stabilization time for 5 dissolved oxygen and temperature AND wait at least 3 seconds before confirming a DO or DO Local calibration Replace membrane and electrolyte 6 Clean sensor...

Страница 23: ...y calibration 3 Verify temperature readings 4 are accurate Sample conductivity is outside 5 the measurement range of the instrument i e 0 200 mS Contact YSI Tech Support 6 S pecificati o ns These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at www ysi com or contact YSI Tech Support...

Страница 24: ... a temperature and conductivity sensor The dissolved oxygen sensor is sold separately D eclarati o n o f C o nf o rmity The undersigned hereby declares on behalf of the named manufacturer under our sole responsibility that the listed product conforms to the requirements for the listed European Council Directive s and carries the CE mark accordingly Manufacturer YSI Incorporated 1725 Brannum Lane Y...

Страница 25: ... and users are expected to understand and follow the battery disposal requirements for their specific locale C o ntact I nf o rmati o n Ordering and Technical Support Telephone 800 897 4151 USA 1 937 767 7241 Globally Monday through Friday 8 00 AM to 5 00 ET Fax 1 937 767 9353 orders 1 937 767 1058 technical support Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs...

Страница 26: ... Air at 760 mm Hg Pressure Salinity Measure of quantity of dissolved salts in water Chlorinity Measure of chloride content by mass of water S 0 00 1 80655 x Chlorinity 0 00 Temp C Chlorinity 0 Salinity 0 5 0 ppt 9 0 ppt 10 0 ppt 18 1 ppt 15 0 ppt 27 1 ppt 20 0 ppt 36 1 ppt 25 0 ppt 45 2 ppt 0 0 14 62 13 73 12 89 12 10 11 36 10 66 1 0 14 22 13 36 12 55 11 78 11 07 10 39 2 0 13 83 13 00 12 22 11 48 ...

Страница 27: ...7 6 73 6 41 6 10 5 81 32 0 7 31 6 96 6 62 6 31 6 01 5 72 33 0 7 18 6 84 6 52 6 21 5 91 5 63 34 0 7 07 6 73 6 42 6 11 5 82 5 55 35 0 6 95 6 62 6 31 6 02 5 73 5 46 36 0 6 84 6 52 6 22 5 93 5 65 5 38 37 0 6 73 6 42 6 12 5 84 5 56 5 31 38 0 6 62 6 32 6 03 5 75 5 48 5 23 39 0 6 52 6 22 5 98 5 66 5 40 5 15 40 0 6 41 6 12 5 84 5 58 5 32 5 08 41 0 6 31 6 03 5 75 5 49 5 24 5 01 42 0 6 21 5 93 5 67 5 41 5 1...

Отзывы: