background image

7

TROUBLESHOOTING 

MAIN 

DISPLAY 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

Er 1 

pH electrode offset is greater/less 
than +/-1.5 pH. 

STAND was pressed before the 
electrode and ATC/Temp probe settle 
to within ±1.5 pH of the buffer value. 

pH electrode is faulty 

Replace the buffer and/or 
the pH electrode. Press ESC.

Allow sufficient time for the 
electrode and ATC/Temp 
probe to stabilize. 

Return for service. 

Er 2 

pH electrode slope is off by more 
than 30% of ideal slope. 

SLOPE was pressed before the 
electrode and ATC/Temp probe 
settled to within 30% of the buffer 
value. 

Buffer 4.00, 4.01, 9.18 or 10.01 is not 
correct. 

Check that the correct buffer 
is used and that the 
electrode slope is not off by 
more than 30% from the 
theoretical slope.

Allow sufficient time for the 
electrode and ATC/Temp 
probe to stabilize. 

Replace the buffer and/or 
the pH electrode. Press ESC. 

Return for service. 

Er 3 

Temperature is out of the 0.0 to 60.0 
°C range. 

Bring the buffer temperature 
within range. 

Return for service.

OvEr/Undr 

Measured pH is out of the 16.00/-2.00 
pH range. 

Measured mV is out of the 1250/-
2000 mV range. 

Measured temperature is out of the
-10/120 °C range. 

Bring the out of range unit 
into the correct measuring 
range.

If units are within proper 
range, return product for 
service. 

SPECIFICATIONS 

Display

Range

Accuracy 

Resolution

pH 

-2.00 to 16.00 pH 

±0.1%, ±2 lsd 

0.01 pH 

mV 

-1999 to 1999 mV

±0.1%, ±1 lsd 

1 mV 

Temperature °C 

-10.0 to 120 °C 

±0.5 °C 

0.1 °C 

pH Temp Compensation 

AUTO/MANual -10.0 to 120.0 

°C

 

pH Buffer Recognition 

(4.01, 7.00 & 10.01) or (4.00, 6.86 & 9.18) 

pH Buffer Calibration Temp. Range 

0 to 60 

°C

 

pH Electrode Offset Recognition 

±90mV at pH 7.00 or 6.86 

pH Electrode Slope Recognition 

±30% at pH 4.00, 4.01, 9.18 or 10.01 

Power  

One 9V battery/ Approximately 1000 hours

 

Instrument Casing 

Waterproof IP 67

Calibration Back-up 

Yes 

Audio Feedback 

Yes, on all touch keys 

Autolock Feature 

Yes 

Operating Temp. Range 

0 to 50 

°C

 

Operating Relative Humidity Range 

up to 95% 

ATC Probe 

Thermistor, 10k

 / 25 

°C

 

Dimensions (L x W x D) 

18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)

 

Weight (batteries included) 

270 grams (.6 lb)

 

Содержание EcoSense pH100A

Страница 1: ...Operations Manual EcoSense pH100A Portable pH mV and Temperature Instrument English ...

Страница 2: ...isposal 3 KEY FUNCTIONS OF THE MODEL pH100A 4 THE LCD DISPLAY 4 OPERATIONAL PROCEDURES 4 Buffer Set Selection 4 pH Calibration 5 Using the model pH160 Electrode Simulator 5 pH Measurements 5 Temperature Measurements 6 mV Measurements 6 Saving Viewing and Deleting Data 6 TROUBLESHOOTING 7 SPECIFICATIONS 7 ...

Страница 3: ...et memory and 50 60 Hz AC noise rejection This meter is universal user friendly for field industrial and laboratory applications INITIAL INSPECTION Carefully unpack the unit and accessories and inspect for shipping damages Compare received parts with materials listed on the packing list Notify YSI immediately of any damage or missing parts Save all packing materials until satisfactory operation is...

Страница 4: ...ternal memory When ESC is pressed for 2 seconds all LCD elements light After about 2 seconds the unit enters pH AUTOLOCK mode AUTOLOCK displays and STAND begins to flash indicating the need for calibration THE LCD DISPLAY 1 WAIT Displays while unit waits for a stable reading or end point sensing 2 BAT Low battery indicator 3 ATC MAN ATC displays if an ATC probe is connected Otherwise MAN displays ...

Страница 5: ... calibrated and ready for measurements After calibration press and hold MEA EFF for about 5 seconds to display the new electrode efficiency Using the model pH160 Electrode Simulator The model pH160 Electrode Simulator can be used to confirm proper instrument calibration To use the simulator 1 Install the 9V battery provided 2 Attach the pH160 to the pH100A Turn both units on The pH160 has a small ...

Страница 6: ...SAVE and the record number for one second Full is displayed when trying to save data and the memory is full To view saved data press mode until RECALL is displayed and then press ESC Use the Up or Down arrow keys to review different saved records Press Mode to escape back to measurement mode To delete data records press Mode while in measurement mode until DELETE is displayed Press ESC All will be...

Страница 7: ...r temperature within range Return for service OvEr Undr Measured pH is out of the 16 00 2 00 pH range Measured mV is out of the 1250 2000 mV range Measured temperature is out of the 10 120 C range Bring the out of range unit into the correct measuring range If units are within proper range return product for service SPECIFICATIONS Display Range Accuracy Resolution pH 2 00 to 16 00 pH 0 1 2 lsd 0 0...

Отзывы: