7

to stabilize if a temperature sensor is connected, or manually 
change the temperature reading with the 

up

 and 

down

 keys if no 

temperature sensor is connected.

5.  Press and hold the 

Stand

 key for 2 seconds to calibrate. The unit 

will immediately be calibrated to the first point. The buffer used for 
the first point will remain lit on the display while the remaining two 
buffers start to flash. The unit is ready to be calibrated using the 
second buffer solution.

•  To complete a one point calibration, press the 

Mode

 key after 

successful calibration using the first buffer.

6.  Rinse the pH and temperature (if applicable) sensors with distilled 

water and immerse in the second buffer solution (either 4.00/4.01 
or 9.18/10.01). Allow sufficient time for the pH measurement 
to stabilize. Also allow the temperature reading to stabilize if 
a temperature sensor is connected, or manually change the 
temperature reading with the 

up

 and 

down

 keys if no temperature 

sensor is connected.

7.  Press the 

Slope

 key to calibrate to the second point. The unit will 

immediately be calibrated to the second point. The buffer used 
for the first and second point will be lit on the display, while the 
remaining buffer will start to flash. The unit is ready to be calibrated 
using the third buffer solution. 

•  To complete a two point calibration, press the 

Mode

 key after 

successful calibration using the first two buffers. 

8.  Repeat steps #5 and #6 for the remaining buffer solution. The 

pH1000A will automatically exit the pH calibration mode after 
successful calibration with the third buffer.

VIEWING THE ELECTRODE EFFICIENCY AFTER CALIBRATION

The pH1000A calculates and compensates for the pH electrode slope 
deviation corresponding to the values of the three calibration buffers. 
After calibration, press and hold 

Meas. / Effic.

 key for 5 seconds to 

display the new electrode efficiency.

Electrode efficiency can be converted into a mV/pH slope value based 
on the following equation, with -59.16 mV/pH being the theoretical 
Nernst Slope:

•  Electrode Efficiency = Observed Slope  * 100

     

-59.16 

mV/pH

A single electrode efficiency will be displayed after a single point 
calibration. This efficiency will always equal 100%, as the theoretical 
Nernst Slope will always be used for the observed slope. This is why 
YSI recommends at least a two point calibration.

Содержание EcoSense pH1000A

Страница 1: ...USER MANUAL English EcoSense pH1000A pH mV ORP and Temperature Benchtop Meter User Manual DOCUMENT 601119REF ...

Страница 2: ...Item 601119REF Rev C October 2017 For the latest version of this manual visit ysi com 2017 YSI Incorporated The YSI logo and EcoSense are registered trademarks of YSI Incorporated ...

Страница 3: ...OCK mode 5 Calibration in the pH NON AUTOLOCK mode 6 Viewing the Electrode Efficiency after Calibration 7 pH Measurements 8 Measurements in the pH AUTOLOCK mode 8 Measurements in the pH NON AUTOLOCK mode 8 Temperature Measurements 9 mV Measurements 9 Measurements in the mV AUTOLOCK mode 9 Measurements in the mV NON AUTOLOCK mode 9 Troubleshooting 10 Specifications 11 pH1000A Instrument Only Specif...

Страница 4: ...Contact Information and Service 12 Ordering and Technical Support 12 Service Information 13 Recycling 13 Appendix A Temperature coefficient of pH buffers 14 ...

Страница 5: ...ucts are warranted for the balance of the original warranty period or at least 90 days from date of repair or replacement LIMITATION OF WARRANTY This Warranty does not apply to any YSI product damage or failure caused by Failure to install operate or use the product in accordance with YSI s written instructions Abuse or misuse of the product Failure to maintain the product in accordance with YSI s...

Страница 6: ...ii T H I S PA G E L E F T I N T E N T I O N A L LY B L A N K ...

Страница 7: ...ree measurements The pH1000A can also be used in non AUTOLOCK mode Other features include single dual or three point calibration electrode offset recognition electrode slope recognition electrode efficiency display built in buffer coefficients automatic or manual temperature compensation and long battery life This meter is user friendly for laboratory application The following sensors are availabl...

Страница 8: ...ts INSTRUMENT POWER The pH1000A can be powered by the power supply included with the instrument or with 6 AA alkaline batteries Check the label on the AC adaptor supplied with the instrument to make sure that the AC line voltage is correct If the wrong AC adaptor is supplied notify your YSI representative immediately INSTALLING THE BATTERIES To install or replace the batteries follow the procedure...

Страница 9: ...ibration Clear Clear key Press to clear any errors that are displayed Press and hold for 2 seconds to clear all calibration values stored in the internal memory After 2 seconds the unit will enter the pH AUTO mode AUTO and CAL will be displayed and one of the buffers in the pre selected buffer set will start to flash This means that the unit must be calibrated again before use Stand Slope is used ...

Страница 10: ... the calibration mode 7 AUTO AUTOLOCK mode indicator 8 Buffer selection This indicator will flash if the pH1000A needs to be recalibrated This indicator also identifies the buffer set currently selected 9 LOCK This will indicate the reading is frozen during AUTOLOCK mode 10 EFF This will be displayed if the user is viewing the efficiency of the electrode It is recommended to clean and or replace t...

Страница 11: ... on the buffer set selected CALIBRATION IN THE pH AUTOLOCK MODE In the pH AUTOLOCK mode the meter will wait for the measurement value to become stable before completing the calibration 1 Turn the unit on Press and hold the Clear key for 2 seconds all LCD elements will light up The meter clears all calibration values stored in internal memory 2 Connect the pH electrode to the BNC input and the temp...

Страница 12: ...rd buffer solution To complete a two point calibration press the Mode key after successful calibration using the first two buffers 7 Repeat steps 5 and 6 for the remaining buffer solution The pH1000A will automatically exit the pH calibration mode after successful calibration with the third buffer CALIBRATION IN THE pH NON AUTOLOCK MODE In the pH NON AUTOLOCK mode the instrument will not wait for ...

Страница 13: ...be calibrated to the second point The buffer used for the first and second point will be lit on the display while the remaining buffer will start to flash The unit is ready to be calibrated using the third buffer solution To complete a two point calibration press the Mode key after successful calibration using the first two buffers 8 Repeat steps 5 and 6 for the remaining buffer solution The pH100...

Страница 14: ...erature can manually be set by pressing the up and down arrow keys 2 Rinse the pH and temperature if applicable sensors with distilled water and immerse in the sample to be measured Remove any air bubbles trapped around the electrode by shaking or stirring the electrode 3 Press the Meas Effic key The WAIT icon will start flashing The unit is waiting for a stable pH and temperature measurement 4 Wh...

Страница 15: ...EASUREMENTS IN THE mV AUTOLOCK MODE 1 Press the Mode key until the mV icon and AUTO icon light up 2 Rinse the electrode with distilled water and immerse in the sample 3 Press the Meas Effic key The WAIT icon will start flashing The unit is waiting for a stable measurement 4 When the WAIT icon disappears and the LOCK icon appears the reading is frozen and will not respond to further changes For sam...

Страница 16: ...fer value 2 Buffer 4 00 4 01 or 9 18 10 01 is not correct 3 pH electrode slope is off by more than 30 of ideal slope 1 Allow sufficient time for the electrode and ATC Temp probe to stabilize re press Slope key to continue the calibration procedure 2 Check if the correct buffer is used 3 Replace the buffer and or the pH electrode Press Clear key to recalibrate meter Er3 1 Temperature is out of the ...

Страница 17: ...C pH1000A GENERAL SPECIFICATIONS Operating Temperature Range 0 C to 50 0 C pH Buffer Recognition pH 7 00 4 01 10 01 or pH 6 86 4 00 9 18 pH Temperature Compensation AUTO MAN 0 0 C to 100 0 C pH Buffer Temperature Range 0 C to 60 0 C pH Electrode Offset Recognition 100 mV at pH 7 00 or 108 3 91 7 mV at pH 6 86 pH Electrode Slope Recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input Impedance 1012 Ω T...

Страница 18: ...ement pH1000A Power Supply C O N TA C T I N F O R M AT I O N A N D S E R V I C E ORDERING AND TECHNICAL SUPPORT Telephone 800 897 4151 US 1 937 767 7241 Globally Monday through Friday 8 00 AM to 5 00 ET Fax 1 937 767 9353 orders 1 937 767 1058 technical support Email info ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs OH 45387 USA Web address ysi com When placing an order please ha...

Страница 19: ... committed to reducing the environmental footprint in the course of doing business Even though materials reduction is the ultimate goal we know there must be a concerted effort to responsibly deal with materials after they ve served a long productive life cycle YSI s recycling program ensures that old equipment is processed in an environmentally friendly way reducing the amount of materials going ...

Страница 20: ...0 o C 4 01 7 11 10 32 4 01 6 98 9 46 5 o C 4 01 7 08 10 25 4 00 6 95 9 39 10 o C 4 00 7 06 10 18 4 00 6 92 9 33 15 o C 4 00 7 03 10 12 4 00 6 90 9 28 20 o C 4 00 7 01 10 06 4 00 6 88 9 23 25 o C 4 01 7 00 10 01 4 00 6 86 9 18 30 o C 4 01 6 98 9 97 4 01 6 85 9 14 35 o C 4 02 6 98 9 93 4 02 6 84 9 10 40 o C 4 03 6 97 9 89 4 03 6 84 9 07 45 o C 4 04 6 97 9 86 4 04 6 83 9 04 50 o C 4 06 6 97 9 83 4 06...

Отзывы: