PROBE OPERATION

  

 

NOTE:

  The YSI 550A DO Instrument should not be used in a purpose other than that specified by 

YSI Incorporated. See 

Warranty 

for details.  

 

STIRRING  

 
It is important to recognize that a very small amount of oxygen dissolved in the sample is consumed 
during probe operation.  It is therefore essential that the sample be continuously stirred at the sensor tip. 
If stagnation occurs, measurements will appear artificially low.  
 
Stirring may be accomplished by mechanically moving the sample around the probe tip, or by moving 
the probe through the sample.  The YSI Model 550A has a flow dependence of <25%.  The rate of 
stirring required is 1/2 foot per second (16cm per second). 
 

MEASUREMENT PROCEDURE 

 
1. 

Insert the probe into the sample to be measured. 

2. 

Continuously stir or move the probe through the sample. 

3. 

Allow temperature and dissolved oxygen readings to stabilize. 

4. 

Observe/Record readings.   

5. 

If possible, rinse the probe with clean water after each use. 

 

PRECAUTIONS  

 
1.  Membranes last longer if properly installed and regularly maintained. Erratic readings can result 

from loose, wrinkled, damaged, or fouled membranes, large (more than 1/8" diameter) air bubbles 
in the electrolyte reservoir, or membrane coating by oxygen consuming (e.g. bacteria) or oxygen 
producing (e.g. algae) organisms.  If unstable readings or membrane damage occurs, replace both 
the membrane cap and electrolyte solution.  

 
2.  Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by behaving like 

oxygen at the probe.  

 
3.  Avoid substances such as acids, caustics, and strong solvents, which may damage probe materials.  

Probe materials include the PE membrane, acrylic plastic, EPR rubber, stainless steel, epoxy, 
polyetherimide and the PVC cable covering. 

 
4.  Always store the probe in the calibration/storage chamber with the moistened sponge. 
 
 

Содержание 550A

Страница 1: ...Operations Manual YSI 550A Dissolved Oxygen Instrument...

Страница 2: ......

Страница 3: ...S OF THE YSI 550A 3 PRINCIPLES OF OPERATION 5 PREPARING THE PROBE 5 DISSOLVED OXYGEN CALIBRATION 6 SALINITY COMPENSATION CALIBRATION 7 PROBE OPERATION 8 PROBE ELECTRODE MAINTENANCE 9 ACCESSORIES REPLA...

Страница 4: ......

Страница 5: ...gle calibration regardless of which dissolved oxygen display is used and will calibrate in either mode Salinity compensation values can be changed at any time without performing a new calibration A de...

Страница 6: ...to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the...

Страница 7: ...and salinity calibration and Increases or decreases the value during calibrations and Mode Press at the same time to switch the temperature units between Fahrenheit F and Celsius C and Mode Press at t...

Страница 8: ...the instrument by unscrewing the two screws 2 Remove the rubber stopper from the chamber by pulling the tail free of the small hole on the chamber 3 Re attached the rubber stopper to the storage chamb...

Страница 9: ...causes the probe current to change proportionally PREPARING THE PROBE MEMBRANE CAP INSTALLATION The YSI 550A DO probe is shipped with a dry protective membrane Before using the instrument for the firs...

Страница 10: ...mation can be obtained over the internet or from a local airport or weather station To convert from meters to feet divide by 0 3048 For best results Check calibration with each use and recalibrate as...

Страница 11: ...ess and release both the UP ARROW and DOWN ARROW keys at the same time to enter the calibration menu 4 Press the Mode key until mg L is displayed on the right side of the screen for oxygen units Press...

Страница 12: ...hrough the sample 3 Allow temperature and dissolved oxygen readings to stabilize 4 Observe Record readings 5 If possible rinse the probe with clean water after each use PRECAUTIONS 1 Membranes last lo...

Страница 13: ...ne does not resolve the problem then proceed with the chemical cleaning Mechanical cleaning Sand off the dark layer from the silver anode with 400 grit wet dry sandpaper Wrap the wet sandpaper around...

Страница 14: ...the hex screw was removed completely make sure that it is not cross threaded when replaced 7 The YSI 559 DO module is shipped with a dry protective membrane Before using for the first time remove the...

Страница 15: ...module Dissolved Oxygen Response Time 95 of end value in 9 seconds Temperature Units Celsius or Fahrenheit user selectable Parameter Compensation Automatic temperature compensation for dissolved oxyge...

Страница 16: ...Return system for service 4 Instrument readings are inaccurate A Verify calibration altitude and salinity settings are correct and recalibrate B Probe may not have been in 100 water saturated air duri...

Страница 17: ...Main display reads Err or burn and Secondary display reads EEP A Return system for service CONTACT INFORMATION YSI offers a wide range of customer assistance and technical support functions to ensure...

Страница 18: ...y from the receiver plug the YSI Instrument into a different outlet so that the computer and receiver are on different branch circuits If necessary the user should consult the dealer or an experienced...

Отзывы: