YSI 5200A
173
YSI 5200A
172
5. Press
to select.
6. Enable and configure submenus. Configure control and alarm relay(s)
using AquaManager - see page 146.
7. Optional - Configure digital 4/20mA output - page 178.
8. Press
to save configuration and return to Run Screen.
9. Test system output device(s) at→System→Service→Output Test - page 141
or create an event to verify configured relay(s) energize.
Notes:
-
It is important to keep the 5200A at the Run Screen during normal opera-
tion. Alarm functionality (email alarms, activation of enabled alarm relays,
alarm icons, and buzzer) is suspended when 5200A is not at Run Screen.
-
One relay can be assigned to multiple 5200A systems. It is essential that the
relay configuration be verified in order to avoid conflict of operation of wired
relay devices. If not configured properly, assigning one relay to multiple sys-
tems can result in invalid operation of 5200A.
-
DO set point, control, and alarm values are always entered in mg/L regardless
of display format.
-
DO set point, control, and alarm values can be up to three (3) characters,
excluding decimal point. Values will truncate to 10ths when 10s digit is
configured.
-
5200A calculates the correct %Sat value based on configured elevation dur-
ing calibration.
-
When DO range low 2 or DO range high 2 are active with energized relays,
only the range low 2 and range high 2 control messages and icons display at
the control menu and Run Screen regardless if DO range low 1 or DO range
high 1 relays are also energized.
-
See 30 Second Sensor Hold Off - page 160.
ORP
Menus →Sensor Setup→ORP
Configure ORP sensor system
Before configuring Conductivity system 1) See - “Use AquaManager to configure
system inputs and outputs” - page 146. AND 2) See Control modes [Set point
Control Mode and Alarms - page 150 and PID control mode - 155.]
1. Connect Probe assembly - 60.
2. Wire relay output device(s) to local instrument, or any networked 5X00
instrument - page 46.
AND/OR
2. Wire relay output device(s) to REM - see IOEM/REM Safety and Instal-
lation guide>Chapter 3>Step 4>Wire REM Relay Outputs.
3. Enter ORP Sensor Setup menu
Menus → System → Sensor Setup → ORP
4. Use
▲
▼
to scroll and highlight submenus.
5. Press
to select.
6. Enable and configure submenus. Configure control and alarm relay(s)
using AquaManager - see page 146.
7. Optional - Configure digital 4/20mA output - page 178.
8. Press
to save configuration and return to Run Screen.
9. Test system output device(s) at→System→Service→Output Test - page 141
or create an event to verify configured relay(s) energize.
Notes:
-
It is important to keep the 5200A at the Run Screen during normal opera-
tion. Alarm functionality (email alarms, activation of enabled alarm relays,
alarm icons, and buzzer) is suspended when 5200A is not at Run Screen.
-
One relay can be assigned to multiple 5200A systems. It is essential that the
relay configuration be verified in order to avoid conflict of operation of wired
relay devices. If not configured properly, assigning one relay to multiple sys-
tems can result in invalid operation of 5200A.
Configuring the 5200A
Configuring the 5200A